Section of Periodontology, Department of Oral Health Sciences, KU Leuven & Dentistry, University Hospitals Leuven, Kapucijnenvoer 33, 3000, Leuven, Belgium.
Clin Oral Investig. 2020 Aug;24(8):2829-2836. doi: 10.1007/s00784-019-03148-9. Epub 2020 Jan 16.
Already decades ago, the diagnosis of halitosis was facilitated with the arrival of chair-side instruments to score the breath odor. These devices are used for a more objective assessment of halitosis compared with organoleptic scoring, but these too have their disadvantages. To overcome some of the drawbacks of the original model of the OralChroma (CHM-1), few years ago a second generation of this machine (CHM-2) was introduced. This study compared both devices in a clinical setting.
All records of the patients visiting a specialized halitosis consultation over a period of 5 years (2012-2017) were examined. The correlations of the OralChroma CHM-1 and CHM-2 with the organoleptic and Halimeter® measurements were analyzed. Additionally the sensitivity, specificity, and negative and positive predicted values were calculated.
A total of 581 data points were included (CHM-1: 292, CHM-2: 289). The correlations between both models with the organoleptic measurements were not statistically significant different. The CHM-2 seemed superior to the CHM-1 in the quantification of dimethyl sulfide with a detection rate of 95% and 61%, respectively. Additionally, the CHM-2 was significantly more sensitive for dimethyl sulfide than the CHM-1. However the latter showed in turn a better sensitivity for hydrogen sulfide and methyl mercaptan.
The CHM-2 showed a better sensitivity for dimethyl sulfide than its predecessor. However, its sensitivity for hydrogen sulfide and methyl mercaptan was worse.
Dimethyl sulfide is the main volatile implicated in extra-oral blood-borne halitosis, this makes the OralChroma CHM-2 the instrument of choice when this is suspected.
几十年前,随着用于评分口气的椅旁仪器的出现,口臭的诊断变得更加便捷。与感官评分相比,这些设备更能客观地评估口臭,但它们也有其缺点。为了克服原始 OralChroma(CHM-1)模型的一些缺陷,几年前推出了该机器的第二代(CHM-2)。本研究在临床环境中对这两种设备进行了比较。
检查了在专门的口臭咨询中就诊的患者的所有记录,时间跨度为 5 年(2012-2017 年)。分析了 OralChroma CHM-1 和 CHM-2 与感官和 Halimeter®测量值的相关性。此外,还计算了灵敏度、特异性、阴性和阳性预测值。
共纳入 581 个数据点(CHM-1:292,CHM-2:289)。两种模型与感官测量值之间的相关性没有统计学意义上的差异。CHM-2 在二甲基硫的定量方面优于 CHM-1,检出率分别为 95%和 61%。此外,CHM-2 对二甲基硫的敏感性明显高于 CHM-1。然而,后者对硫化氢和甲硫醇的敏感性更好。
CHM-2 对二甲基硫的敏感性优于其前身。然而,其对硫化氢和甲硫醇的敏感性较差。
二甲基硫是与外源性血源性口臭相关的主要挥发性物质,这使得当怀疑有这种情况时,OralChroma CHM-2 成为首选仪器。