Real time sound spectral analysis for diagnosis of thrombosed prosthetic valves.

作者信息

Sato N, Mohri H, Kagawa Y, Horiuchi T, Tanaka M, Nitta S

机构信息

Department of Thoracic and Cardiovascular Surgery, Tohoku University School of Medicine, Sendai, Japan.

出版信息

ASAIO Trans. 1988 Jul-Sep;34(3):831-4.

PMID:3196607
Abstract

A real time sound spectral analyzer was developed at the authors' institute and has been used for diagnosis of thrombosed valves for 15 years. The analyzer consists of a wider frequency microphone, a digital signal processor with multiple bandpass filters, and displays. To quantify the high frequency component in section pattern of opening and closing clicks, maximal frequency at -30 dB level was used at a diagnostic parameter (normalized maximal frequency, NMF). A total of 17 thrombosed valves with or without pannus formations have been experienced and SSA showed abnormality in 13 valves, which comprised of Björk-Shiley (N = 6), Starr-Edwards 2320, 6120, 6520 (N = 5) and Kay-Shiley (N = 2). The abnormal findings included NMF of opening or closing clicks lower than normal range of each type of prosthesis, and significant decrease of NMF more than 450 Hz from the value previously obtained. All diagnoses of thrombus formations and their locations were confirmed by operative or autopsy findings. The SSA failed to diagnose in four valves, which comprised of Starr-Edwards 6520 (N = 3) and Björk-Shiley (N = 1). Operative findings revealed that thrombi were formed in a manner that permitted direct collision of disc and valvular ring causing clicks with high frequency component. No pseudo-positive diagnosis for thrombosed valve was experienced in the authors series, and SSA is thought to be a reliable noninvasive technique that enables early diagnosis of a thrombosed valve.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验