使用泛影葡胺提高锥形束 CT 图像诊断牙隐裂的准确性。

Using Meglumine Diatrizoate to improve the accuracy of diagnosis of cracked teeth on Cone-beam CT images.

机构信息

Department of Dentomaxillofacial Radiology, Nanjing Stomatological Hospital, Medical School of Nanjing University, Nanjing, China.

Department of Stomatology, Third People's Hospital of Danyang City, Danyang, China.

出版信息

Int Endod J. 2020 May;53(5):709-714. doi: 10.1111/iej.13270. Epub 2020 Feb 17.

Abstract

AIM

To explore in a laboratory setting the feasibility of using Meglumine Diatrizoate (MD) to improve the accuracy of diagnosis of cracked teeth on cone-beam CT (CBCT) images.

METHODOLOGY

Twenty-four teeth were cracked artificially by soaking them cyclically in liquid nitrogen and hot water. The number and position of crack lines were evaluated with a dental operating microscope and used as the gold standard. The artificially cracked teeth were then examined using routine scanning (RS) and enhanced scanning (ES) modes, respectively. For the ES mode, MD was painted on the surface of the crack lines, and then, CBCT scanning with the same parameters was performed after 10 min. A radiological graduate student and an experienced radiologist evaluated the presence or absence of crack lines, respectively. The differences between the RS and ES modes were determined and assessed using McNemar's test. Inter-examiner agreement and intra-examiner agreement were assessed using kappa analysis.

RESULTS

Fifty-seven crack lines were found in the 24 cracked teeth. In the RS mode, the accuracy of detection of crack lines was 23% (radiological graduate student) and 32% (experienced radiologist), whereas in the ES mode, the accuracy was 61% (radiological graduate student) and 65% (experienced radiologist). The inter-examiner agreement was 0.693 in RS mode and 0.849 in ES mode. The intra-examiner agreement was 0.872 and 0.949 for the radiological graduate student in RS and ES mode respectively; and one for the experienced radiologist both in RS and ES mode.

CONCLUSIONS

Compared with routine scanning mode, more crack lines could be detected in enhanced scanning mode using Meglumine Diatrizoate as a contrast medium. MD could be a potential contrast medium to improve the accuracy of detection of crack lines on CBCT images.

摘要

目的

在实验室环境中探索使用泛影葡胺(MD)提高 CBCT 图像上诊断牙隐裂准确性的可行性。

方法

将 24 颗牙齿浸泡在液氮和热水中进行周期性循环,使其人为开裂。用牙科手术显微镜评估裂纹线的数量和位置,并将其作为金标准。然后分别用常规扫描(RS)和增强扫描(ES)模式检查人为开裂的牙齿。对于 ES 模式,将 MD 涂在裂纹线上,然后在 10 分钟后用相同参数进行 CBCT 扫描。一名放射科研究生和一名经验丰富的放射科医生分别评估裂纹线的存在或不存在。使用 McNemar 检验确定 RS 和 ES 模式之间的差异,并进行评估。使用 Kappa 分析评估两位检查者之间的一致性和同一位检查者的重复性。

结果

在 24 颗开裂的牙齿中发现了 57 条裂纹线。在 RS 模式下,检测裂纹线的准确性分别为研究生 23%和经验丰富的放射科医生 32%,而在 ES 模式下,准确性分别为研究生 61%和经验丰富的放射科医生 65%。RS 模式下两位检查者之间的一致性为 0.693,ES 模式下为 0.849。在 RS 模式下,研究生的内部一致性分别为 0.872 和 0.949,而经验丰富的放射科医生在 RS 和 ES 模式下均为 1.00。

结论

与常规扫描模式相比,使用泛影葡胺作为造影剂的增强扫描模式可以检测到更多的裂纹线。MD 可能是一种潜在的造影剂,可提高 CBCT 图像上裂纹线检测的准确性。