Prone Myocardial Perfusion Imaging and Breast Attenuation: A Phantom Study.

作者信息

Ramos Susie Medeiros Oliveira, Glavam Adriana Pereira, de Brito Adriana Soares Xavier, Kubo Tadeu Takao Almodovar, Tukamoto Gustavo, Sampaio Debora da Cunha Pereira, de Sá Lidia Vasconcellos

机构信息

Radiology Department, Federal University of Rio de Janeiro (UFRJ), Clementino Fraga Filho University Hospital, Rio de Janeiro, Brazil.

Nuclear Medicine Department, Clínica de Diagnóstico Por Imagem - Diagnósticos da América SA (CDPI/DASA), Rio de Janeiro, Brazil.

出版信息

Curr Med Imaging Rev. 2020;16(1):70-79. doi: 10.2174/1573405614666181017121453.

Abstract

BACKGROUND

Soft tissue attenuation artifacts are the most common cause of misinterpretation in myocardial perfusion Imaging (MPI). Few studies assessing the value of prone imaging in women have been published. Breast attenuation artifacts can be present in up to 40% of the MPI studies in women.

OBJECTIVES

This study aimed at evaluating the potential impact of prone MPI on breast attenuation, with a critical analysis of activity optimization and breast size influence.

METHODS

MPI of an Anthropomorphic Torso Phantom with silicone breast prostheses and equivalent adipose tissue was compared to a standard MPI database.

RESULTS

A medical qualitative and semiquantitative analysis demonstrated higher uptake in the LV anterior segments in the prone position for all injected activities. An artificial myocardium lesion was diagnosable in the right segment in all images, which shows that prone positioning would not mask a true lesion and it assists the cardiologist with a more accurate analysis. These results showed that it is possible to optimize the activity to be injected by up to 55.6% when using combined supine-prone images.

CONCLUSION

Prone position has a high impact on the interpretation of MPI in female patients since it reduces the breast attenuation artifacts, and optimizes the radiation protection of the patient and all staff involved in the procedure, making it more cost-effective.

摘要

背景

软组织衰减伪像是心肌灌注成像(MPI)中最常见的误判原因。很少有评估俯卧位成像对女性价值的研究发表。在女性的MPI研究中,高达40%可能存在乳腺衰减伪像。

目的

本研究旨在评估俯卧位MPI对乳腺衰减的潜在影响,并对活性优化和乳房大小的影响进行批判性分析。

方法

将带有硅胶乳房假体和等效脂肪组织的拟人化躯干模型的MPI与标准MPI数据库进行比较。

结果

医学定性和半定量分析表明,在所有注入活性的情况下,俯卧位时左心室前壁节段的摄取更高。在所有图像中,右节段均可诊断出人工心肌病变,这表明俯卧位不会掩盖真正的病变,有助于心脏病专家进行更准确的分析。这些结果表明,使用仰卧位与俯卧位联合图像时,可将注入活性优化高达55.6%。

结论

俯卧位对女性患者MPI的解读有很大影响,因为它减少了乳腺衰减伪像,优化了患者及参与该程序的所有工作人员的辐射防护,使其更具成本效益。