Aleksa Gytis, Jaruševičius Paulius, Pacaitytė Andrė, Vajauskas Donatas
Radiology Clinic, Medical Academy, Faculty of Medicine, Lithuanian University of Health Sciences, 44307 Kaunas, Lithuania.
Department of Nuclear Medicine, Radiology Clinic, Lithuanian University of Health Sciences, Kaunas Clinics, 44307 Kaunas, Lithuania.
Medicina (Kaunas). 2025 Apr 4;61(4):665. doi: 10.3390/medicina61040665.
Single-photon emission computed tomography (SPECT) myocardial perfusion imaging (MPI) is a well-established technique for evaluating myocardial perfusion and function in patients with suspected or known coronary artery disease. While conventional dual-detector SPECT scanners have limitations in spatial resolution and photon detection sensitivity, recent advancements, including full-ring solid-state cadmium zinc telluride (CZT) detectors, offer enhanced image quality and improved diagnostic accuracy. This study aimed to compare the performance of Veriton-CT, a full-ring CZT SPECT system, with GE Discovery 530c, a dedicated cardiac fixed-angle gamma camera, in myocardial perfusion imaging and their correlation with coronary angiography findings. This was a prospective study that analyzed 21 patients who underwent MPI at the Department of Nuclear Medicine, Lithuanian University of Health Sciences, Kauno Klinikos. A one-day stress-rest protocol using 99mTc-Sestamibi was employed, with stress testing performed via bicycle ergometry or pharmacological induction. MPI was first conducted using GE Discovery 530c (GE Health Care, Boston, MA, USA), followed by imaging on Veriton-CT, which included low-dose CT for attenuation correction. The summed stress score (SSS), summed rest score (SRS), and summed difference score (SDS) were analyzed and compared between both imaging modalities. Coronary angiography results were retrospectively collected, and lesion-based analysis was performed to assess the correlation between imaging results and the presence of significant coronary artery stenosis (≥35% and ≥70% narrowing). Image quality and the certainty of distinguishing the inferior myocardial wall from extracardiac structures were also evaluated by two independent researchers with differing levels of experience. : Among the 14 patients included in the final analysis, Veriton-CT was more likely to classify MPI scans as normal (64.3%) compared to GE Discovery 530c (28.6%). Additionally, Veriton-CT provided a better assessment of the right coronary artery (RCA) basin, showing greater agreement with coronary angiography findings than GE Discovery 530c, although the difference was not statistically significant. No significant differences in lesion overlap were observed for the left anterior descending artery (LAD) or left circumflex artery (LCx) basins. Furthermore, the image quality assessment revealed slightly better delineation of extracardiac structures using Veriton-CT (Spectrum Dynamics Medical, Caesarea, Israel), particularly when evaluated by an experienced researcher. However, no significant difference was observed when assessed by a less experienced observer. Our findings suggest that Veriton-CT, with its full-ring CZT detector system, may offer advantages over fixed-angle gamma cameras in improving image quality and reducing attenuation artifacts in MPI. Although the difference in correlations with coronary angiography findings was not statistically significant, Veriton-CT showed a trend toward better agreement, particularly in the RCA basin. These results indicate that full-ring SPECT imaging could improve the diagnostic accuracy of non-invasive MPI, potentially reducing the need for unnecessary invasive angiography. Further studies with larger patient cohorts are required to confirm these findings and evaluate the clinical impact of full-ring SPECT technology in myocardial perfusion imaging.
单光子发射计算机断层扫描(SPECT)心肌灌注成像(MPI)是一种成熟的技术,用于评估疑似或已知冠状动脉疾病患者的心肌灌注和功能。虽然传统的双探测器SPECT扫描仪在空间分辨率和光子检测灵敏度方面存在局限性,但包括全环固态碲化镉锌(CZT)探测器在内的最新进展提高了图像质量并提升了诊断准确性。本研究旨在比较全环CZT SPECT系统Veriton-CT与专用心脏固定角度伽马相机GE Discovery 530c在心肌灌注成像中的性能,以及它们与冠状动脉造影结果的相关性。这是一项前瞻性研究,分析了立陶宛卫生科学大学考纳斯临床医院核医学科接受MPI检查的21例患者。采用了使用99mTc-甲氧基异丁基异腈的一日负荷-静息方案,负荷试验通过踏车运动或药物诱导进行。首先使用GE Discovery 530c(美国马萨诸塞州波士顿市GE医疗集团)进行MPI检查,随后在Veriton-CT上成像,其中包括用于衰减校正的低剂量CT。分析并比较了两种成像方式的负荷总分(SSS)、静息总分(SRS)和差值总分(SDS)。回顾性收集冠状动脉造影结果,并进行基于病变的分析,以评估成像结果与显著冠状动脉狭窄(≥35%和≥70%狭窄)的存在之间的相关性。两名经验水平不同的独立研究人员还对图像质量以及区分心肌下壁与心外结构的确定性进行了评估。在最终分析纳入的14例患者中,与GE Discovery 530c(28.6%)相比,Veriton-CT更有可能将MPI扫描分类为正常(64.3%)。此外,Veriton-CT对右冠状动脉(RCA)区域的评估更好,与冠状动脉造影结果的一致性比GE Discovery 530c更高,尽管差异无统计学意义。在左前降支(LAD)或左旋支(LCx)区域,未观察到病变重叠的显著差异。此外,图像质量评估显示,使用Veriton-CT(以色列凯撒利亚市Spectrum Dynamics Medical公司)能更好地勾勒心外结构,尤其是由经验丰富的研究人员评估时。然而,由经验较少的观察者评估时未观察到显著差异。我们的研究结果表明,配备全环CZT探测器系统的Veriton-CT在改善MPI图像质量和减少衰减伪影方面可能优于固定角度伽马相机。尽管与冠状动脉造影结果的相关性差异无统计学意义,但Veriton-CT显示出更好一致性的趋势,尤其是在RCA区域。这些结果表明,全环SPECT成像可提高无创MPI的诊断准确性,可能减少不必要的有创血管造影需求。需要对更大患者队列进行进一步研究以证实这些发现,并评估全环SPECT技术在心肌灌注成像中的临床影响。