Comparison of strain and dyssynchrony measurements in fetal two-dimensional speckle tracking echocardiography using Philips and TomTec.

机构信息

Department of Obstetrics and Gynecology, University Hospital of Münster, Münster, Germany.

出版信息

J Perinat Med. 2020 Mar 26;48(3):266-273. doi: 10.1515/jpm-2019-0424.

Abstract

Background The aim of this study was to compare Philips and TomTec two-dimensional speckle tracking echocardiography (2D-STE) software measurements of strain and dyssynchrony values in healthy fetuses. Methods This was an explorative observational study in which the echocardiographic data of 93 healthy fetuses between the 20th and 38th week of gestation were determined from a four-chamber view using 2D speckle tracking. The global and segmental longitudinal strain values of both ventricles, inter-ventricular and left intra-ventricular dyssynchrony were analyzed using QLab version 10.8 (Philips Medical Systems, Andover, MA, USA) and TomTec-Arena version 2.30 (TomTec, Unterschleißheim, Germany). Results TomTec showed persistently lower values for all of the assessed strain and dyssynchrony variables. For all variables, the bias between vendors tended to increase with gestational age, though not to a significant extent. Left ventricular dyssynchrony and longitudinal strain within the mid segment of the septum correlated best between vendors; however, the limits of agreement were wide in both cases. None of the variables assessed in the two-chamber view compared well between QLAB and TomTec. Conclusion Speckle tracking software cannot be used interchangeably between vendors. Further investigations are necessary to standardize fetal 2D-STE.

摘要

背景 本研究旨在比较飞利浦和 TomTec 二维斑点追踪超声心动图(2D-STE)软件在健康胎儿应变和不同步值测量方面的差异。

方法 这是一项探索性观察性研究,使用二维斑点追踪技术从四腔心切面获取 93 例 20 至 38 孕周健康胎儿的超声心动图数据。使用 QLab 版本 10.8(飞利浦医疗系统,安多弗,MA,美国)和 TomTec-Arena 版本 2.30(TomTec,乌特施莱歇尔海姆,德国)分析心室的整体和节段性纵向应变值以及室间隔和左室内不同步值。

结果 TomTec 对所有评估的应变和不同步变量均显示出持续较低的值。尽管在一定程度上,但随着胎龄的增加,供应商之间的偏差趋势增加。尽管在这两种情况下,协议的范围都很宽,但供应商之间的间隔中隔中部的左心室不同步和纵向应变相关性最佳。在两腔心视图中评估的变量均无法在 QLAB 和 TomTec 之间很好地比较。

结论 斑点追踪软件不能在供应商之间互换使用。需要进一步的研究来规范胎儿 2D-STE。