Copeland K R, Yatscoff R W
Department of Clinical Chemistry, Health Sciences Centre, Winnipeg, Manitoba, Canada.
Ther Drug Monit. 1988;10(4):453-8. doi: 10.1097/00007691-198804000-00013.
The evaluation of a radioimmunoassay (RIA) for the monitoring of cyclosporine (CSA) that utilizes a monoclonal antibody specific for the parent compound is described. The assay exhibited good analytical recovery (greater than 95%), precision (coefficient of variation less than 10%), and linearity in plasma and whole blood specimens. The percentage cross reactivity of the major CSA metabolites M-17, M-1, and M-21 was less than 2%. The RIA gave values that were on average indistinguishable from those obtained by high-pressure liquid chromatography in plasma and whole blood specimens obtained from 49 renal transplant recipients.
本文描述了一种利用针对母体化合物的单克隆抗体监测环孢素(CSA)的放射免疫分析(RIA)方法。该分析方法在血浆和全血样本中表现出良好的分析回收率(大于95%)、精密度(变异系数小于10%)和线性。主要CSA代谢物M-17、M-1和M-21的交叉反应率小于2%。在49名肾移植受者的血浆和全血样本中,该RIA所得值与高压液相色谱法所得值平均无显著差异。