Yatscoff R W, Copeland K R, Faraci C J
Department of Clinical Chemistry, Health Sciences Clinical Research Centre, Winnipeg, Manitoba, Canada.
Clin Chem. 1990 Nov;36(11):1969-73.
We report here the evaluation of the Abbott TDx assay with a monoclonal antibody for selectively quantifying cyclosporine (CsA) in whole blood. Over the clinically relevant concentration ranges, results with this assay demonstrated within- and between-run CVs of less than 2.5% and 5%, respectively; sensitivity of 25 micrograms/L; good analytical recovery (100.3%); and linearity with whole-blood specimens. The percentage cross-reactivity of the major CsA metabolites varied from 15.3% for AM9 (M-1), 8.2% for AM1 (M-17), and 3.7% for AM4N (M-21), to less than 3% for the other metabolites tested. Results with the TDx assay (y) correlated well with those by the Sandimmune selective RIA (x; Sandoz) with blood specimens from 44 renal-transplant recipients (n = 44, x= 187.3, y = 198.9, y = 5.49 + 1.03x, r = 0.987). The TDx values were on average 24% higher than those by HPLC (x') with the same patients' specimens (n = 44, x' = 159.9, y = 198.9, y = 15.9 + 1.14x', r = 0.967). We conclude that the Abbott TDx monoclonal antibody assay provides a rapid, precise, and accurate means for quantifying CsA in whole blood.
我们在此报告使用单克隆抗体的雅培TDx检测法对全血中环孢素(CsA)进行选择性定量的评估。在临床相关浓度范围内,该检测法的批内和批间变异系数分别小于2.5%和5%;灵敏度为25微克/升;分析回收率良好(100.3%);并且与全血标本具有线性关系。主要CsA代谢物的交叉反应百分比从AM9(M - 1)的15.3%、AM1(M - 17)的8.2%、AM4N(M - 21)的3.7%到其他测试代谢物的小于3%不等。对于44名肾移植受者的血液标本,TDx检测法的结果(y)与山德士的环孢素选择性放射免疫分析法(x)的结果相关性良好(n = 44,x = 187.3,y = 198.9,y = 5.49 + 1.03x,r = 0.987)。使用相同患者的标本时,TDx值平均比高效液相色谱法(x')的结果高24%(n = 44,x' = 159.9,y = 198.9,y = 15.9 + 1.14x',r = 0.967)。我们得出结论,雅培TDx单克隆抗体检测法为全血中CsA的定量提供了一种快速、精确且准确的方法。