Department of Radiology and Nuclear Medicine, Radboud University Medical Center, P.O. Box 9101 (Route 766), 6500 HB, Nijmegen, The Netherlands.
Canon Medical Systems Corporation, Otawara, Japan.
Eur Radiol. 2020 May;30(5):2552-2560. doi: 10.1007/s00330-019-06635-5. Epub 2020 Feb 10.
To evaluate the technical performance of an ultra-high-resolution CT (UHRCT) system.
The physico-technical capabilities of a novel commercial UHRCT system were assessed and compared with those of a current-generation multi-detector (MDCT) system. The super-high-resolution (SHR) mode of the system uses 0.25 mm (at isocentre) detector elements (dels) in the in-plane and longitudinal directions, while the high-resolution (HR) mode bins two dels in the longitudinal direction. The normal-resolution (NR) mode bins dels 2 × 2, resulting in a del-size equivalent to that of the MDCT system. In general, standard procedures and phantoms were used to perform these assessments.
The UHRCT MTF (10% MTF 4.1 lp/mm) is twice as high as that of the MDCT (10% MTF 1.9 lp/mm), which is comparable to the MTF in the NR mode (10% MTF 1.7 lp/mm). The width of the slice sensitivity profile in the SHR mode (FWHM 0.45 mm) is about 60% of that of the MDCT (FWHM 0.77 mm). Uniformity and CT numbers are within the expected range. Noise in the high-resolution modes has a higher magnitude and higher frequency components compared with MDCT. Low-contrast visibility is lower for the NR, HR and SHR modes compared with MDCT, but about a 14%, for NR, and 23%, for HR and SHR, dose increase gives the same results.
HR and SHR mode scanning results in double the spatial resolution, with about a 23% increase in dose required to achieve the same low-contrast detectability.
• Resolution on UHRCT is up to twice as high as for the tested MDCT. • With abdominal settings, UHRCT needs higher dose for the same low-contrast detectability as MDCT, but dose is still below achievable levels as defined by current diagnostic reference levels. • The UHRCT system used in normal-resolution mode yields comparable resolution and noise characteristics as the MDCT system.
评估超高清 CT(UHRCT)系统的技术性能。
评估新型商用 UHRCT 系统的物理技术性能,并将其与当前一代多探测器(MDCT)系统进行比较。该系统的超高分辨率(SHR)模式在平面内和纵向方向上使用 0.25 毫米(在等中心处)探测器元件(dels),而高分辨率(HR)模式则在纵向方向上将两个 dels 合并在一起。正常分辨率(NR)模式将 dels 合并为 2×2,从而得到与 MDCT 系统相当的 del 尺寸。通常,使用标准程序和体模进行这些评估。
UHRCT 的调制传递函数(10% MTF 4.1 lp/mm)是 MDCT(10% MTF 1.9 lp/mm)的两倍,与 NR 模式的调制传递函数(10% MTF 1.7 lp/mm)相当。SHR 模式下的切片灵敏度分布宽度(半峰全宽 0.45 毫米)约为 MDCT(半峰全宽 0.77 毫米)的 60%。均匀性和 CT 数在预期范围内。高分辨率模式下的噪声比 MDCT 具有更高的幅度和更高的频率成分。与 MDCT 相比,NR、HR 和 SHR 模式的低对比度可见度较低,但 NR 约为 14%,HR 和 SHR 约为 23%,增加剂量会产生相同的结果。
HR 和 SHR 模式扫描结果的空间分辨率提高了一倍,达到相同的低对比度检测能力需要增加约 23%的剂量。
· UHRCT 的分辨率高达 MDCT 的两倍。
· 在腹部设置下,UHRCT 达到与 MDCT 相同的低对比度检测能力需要更高的剂量,但仍低于当前诊断参考水平定义的可达到的剂量水平。
· 使用正常分辨率模式的 UHRCT 系统产生的分辨率和噪声特性与 MDCT 系统相当。