Keizer Michèle N J, Otten Egbert
University of Groningen, University Medical Center Groningen, Center of Human Movement Sciences, Antonius deusinglaan, Groningen, 9713 AV, The Netherlands.
J Exp Orthop. 2020 Feb 10;7(1):6. doi: 10.1186/s40634-020-0219-z.
The first aim was to report the sensitivity of calculated tibiofemoral movements for the choice of placement of the set of femoral markers. The second aim was to report the influence of accuracy of the motion captured positions of the markers on the calculated tibiofemoral movements.
Tibiofemoral kinematics during single leg hops for distance were calculated. For the first aim, an experiment was conducted in which four different setups of the femoral markers were used to calculated tibiofemoral movements. For the second aim, an experiment was conducted in which all raw marker positions were mathematically moved independently with the known Vicon position error with a distance and in a random direction in each frame, repeated a hundred times. Each time, the tibiofemoral movements were calculated.
The first experiment yields that the standard deviation of the calculated anterior tibia translation between marker setups was 0.88 mm and the standard deviation of the external tibia rotation between marker setups was 0.76 degrees. The second experiment yields that the standard deviation was 0.76 mm for anterior tibia translation and 0.38 degrees for external tibia rotation.
A combined standard deviation of both experiments revealed that transients in anterior tibia translation less than 2.32 mm and external tibia rotations less than 1.70 degrees should be taken with caution. These results are 19.42% of the range of the anterior tibia translation and 13.51% of the rotation range during the jump task. The marker setup should be chosen carefully.
第一个目的是报告计算得到的胫股关节运动对于股骨标记点组放置选择的敏感性。第二个目的是报告标记点运动捕捉位置的准确性对计算得到的胫股关节运动的影响。
计算单腿跳远过程中的胫股关节运动学数据。对于第一个目的,进行了一项实验,其中使用四种不同的股骨标记点设置来计算胫股关节运动。对于第二个目的,进行了一项实验,其中所有原始标记点位置在每一帧中都以已知的Vicon位置误差在距离和随机方向上独立进行数学移动,重复一百次。每次都计算胫股关节运动。
第一个实验得出,标记点设置之间计算得到的胫骨前移标准差为0.88毫米,标记点设置之间胫骨外旋标准差为0.76度。第二个实验得出,胫骨前移标准差为0.76毫米,胫骨外旋标准差为0.38度。
两个实验的综合标准差表明,胫骨前移小于2.32毫米和胫骨外旋小于1.70度的瞬态应谨慎对待。这些结果分别是跳跃任务中胫骨前移范围的19.42%和旋转范围的13.51%。应仔细选择标记点设置。