澳大利亚成人CT的更新诊断参考水平。

Updated Australian diagnostic reference levels for adult CT.

作者信息

Lee Kam L, Beveridge Toby, Sanagou Masoumeh, Thomas Peter

机构信息

Australian Radiation Protection and Nuclear Safety Agency, Yallambie, Victoria, Australia.

出版信息

J Med Radiat Sci. 2020 Mar;67(1):5-15. doi: 10.1002/jmrs.372. Epub 2020 Feb 10.

Abstract

INTRODUCTION

In 2018, ARPANSA published updated national DRLs for adult CT, which were first published in 2012, and augmented the national DRL categories. This paper presents the updated national DRLs and describes the process by which they were produced.

METHODS

Examine patient survey data submitted to the Australian Radiation Protection and Nuclear Safety Agency (ARPANSA) National Diagnostic Reference Level Service (NDRLS). Determine the quartiles of the distributions of median survey dose metrics with categorisation by procedure type. Engage a liaison panel representing the radiology professions to review procedure categories and recommend revised national DRLs. The revised NDRL procedure categories are: head (non-contrast brain (trauma/headache)), cervical spine (Non-contrast (trauma)), soft-tissue neck (post-contrast (oncology)), chest (post-contrast (oncology)), abdomen-pelvis (post-contrast (oncology)), kidney-ureter-bladder (non-contrast (suspected renal colic)), chest-abdomen-pelvis (post-contrast (oncology)) and lumbar spine (non-contrast (degenerative pain)).

RESULTS

The existing six procedure categories were revised and refined. Updated Australian national diagnostic reference levels for adult CT were recommended and endorsed for eight procedure categories: head (52 mGy/880 mGycm), cervical spine (23 mGy/470 mGycm),soft-tissue neck (17 mGy/450 mGycm), chest (10 mGy/390 mGycm), abdomen-pelvis (13 mGy/600 mGycm), kidney-ureter-bladder (13 mGy/600 mGycm), chest-abdomen-pelvis (11 mGy/940 mGycm) and lumbar spine (26 mGy/670 mGycm). The updated national DRLs are between 12 and 26% lower than the previous DRLs for dose-length product and between 13 and 63% lower for volume computed tomography dose index.

CONCLUSIONS

Australian national DRLs for adult CT have been reviewed and revised. The revised national DRLs are lower, better reflecting current practice among imaging facilities in Australia. The revised Australian national DRLs are similar to those in other developed countries.

摘要

引言

2018年,澳大利亚辐射防护与核安全局(ARPANSA)发布了成人CT的更新版国家诊断参考水平(DRLs),该水平于2012年首次发布,并扩充了国家DRL类别。本文介绍了更新后的国家DRLs,并描述了其制定过程。

方法

审查提交给澳大利亚辐射防护与核安全局(ARPANSA)国家诊断参考水平服务(NDRLS)的患者调查数据。确定按检查类型分类的调查剂量中位数分布的四分位数。聘请一个代表放射专业的联络小组审查检查类别,并推荐修订后的国家DRLs。修订后的NDRL检查类别为:头部(非增强脑(创伤/头痛))、颈椎(非增强(创伤))、颈部软组织(增强后(肿瘤学))、胸部(增强后(肿瘤学))、腹部-盆腔(增强后(肿瘤学))、肾-输尿管-膀胱(非增强(疑似肾绞痛))、胸部-腹部-盆腔(增强后(肿瘤学))和腰椎(非增强(退行性疼痛))。

结果

现有的六个检查类别得到了修订和完善。推荐并认可了成人CT的八个检查类别的更新后的澳大利亚国家诊断参考水平:头部(52 mGy/880 mGycm)、颈椎(23 mGy/470 mGycm)、颈部软组织(17 mGy/450 mGycm)、胸部(10 mGy/390 mGycm)、腹部-盆腔(13 mGy/600 mGycm)、肾-输尿管-膀胱(13 mGy/600 mGycm)、胸部-腹部-盆腔(11 mGy/940 mGycm)和腰椎(26 mGy/670 mGycm)。更新后的国家DRLs在剂量长度乘积方面比之前的DRLs低12%至26%,在容积CT剂量指数方面低13%至63%。

结论

澳大利亚成人CT的国家DRLs已得到审查和修订。修订后的国家DRLs更低,更好地反映了澳大利亚影像机构的当前实践。修订后的澳大利亚国家DRLs与其他发达国家的类似。