Eur J Radiol. 2020 Apr;125:108863. doi: 10.1016/j.ejrad.2020.108863. Epub 2020 Feb 6.
To define National Diagnostic Reference Levels (NDRLs) for computed tomography (CT) for Austria on adult patients.
In the course of a nationwide survey on common CT-examinations between June 2018 and November 2018, datasets were requested from all Austrian hospitals and radiology centers with CT-scanners. All facilities were asked to report a minimum sample of 10 representative dose length product (DLP) values per examination type and CT-scanner, including information about scan sequences, iterative reconstruction (IR) algorithms and, if available, patient data. Examination types included into the survey were the most common CT-indications for the anatomic regions head, facial bones, thorax, abdomen and pelvis. 3rd quartiles of CT-scanner median DLPs were calculated and compared to Austrian and international NDRLs.
For 76 % of all operating CT-scanners, DLP data was submitted varying from 13 to 172 scanners per examination type. Mean year of manufacture of the CT-scanners was 2011, ranging from 2003 to 2018. In 66 % of the examinations, IR was used. 3rd quartile is on average 21 % lower for scanners using IR algorithms as opposed to scanners or protocols not using IR. In the case of gender differences, the DLP for females is on average 17 % lower than for males.
The new recommendations for CT NDRLs for Austria based on a nationwide survey are on average 13 % lower than the current Austrian NDRLs and correspond well to recent German and Swiss NDRLs. 3rd quartiles for DLP are dependent on gender, the use of IR and year of manufacture of the CT-scanners.
为奥地利成年人制定计算机断层扫描(CT)的国家诊断参考水平(NDRL)。
在 2018 年 6 月至 11 月期间进行的一项全国性 CT 检查调查中,要求奥地利所有拥有 CT 扫描仪的医院和放射科中心提供数据集。所有机构都被要求报告每个检查类型和 CT 扫描仪的至少 10 个具有代表性的剂量长度乘积(DLP)值,包括扫描序列、迭代重建(IR)算法的信息,如果有的话,还包括患者数据。调查中包括的检查类型是头部、面部骨骼、胸部、腹部和骨盆等解剖区域最常见的 CT 适应症。计算 CT 扫描仪中位数 DLP 的第 3 四分位数,并与奥地利和国际 NDRL 进行比较。
对于所有运行的 CT 扫描仪,有 76%提交了 DLP 数据,每个检查类型的 DLP 数据从 13 到 172 个不等。CT 扫描仪的平均制造年份为 2011 年,范围从 2003 年到 2018 年。在 66%的检查中,使用了 IR。与不使用 IR 算法的扫描仪或协议相比,使用 IR 算法的扫描仪的第 3 四分位数平均低 21%。在性别差异方面,女性的 DLP 平均比男性低 17%。
基于全国性调查的奥地利 CT NDRL 的新建议比当前的奥地利 NDRL 平均低 13%,与最近的德国和瑞士 NDRL 非常吻合。DLP 的第 3 四分位数取决于性别、IR 的使用和 CT 扫描仪的制造年份。