首个人体连续 24 小时眼内压和眼搏动的新型隐形眼镜传感器测量。

First-in-human continuous 24-hour measurement of intraocular pressure and ocular pulsation using a novel contact lens sensor.

机构信息

[WASILEWICZ]-Eye Clinic, Poznan, Poland.

Sensimed AG, Lausanne, Switzerland.

出版信息

Br J Ophthalmol. 2020 Nov;104(11):1519-1523. doi: 10.1136/bjophthalmol-2019-315276. Epub 2020 Feb 19.

Abstract

BACKGROUND/AIMS: This study assessed the feasibility of a novel contact lens device for intraocular pressure (IOP) and ocular pulse amplitude (OPA) continuous measurements over 24 hours.

METHODS

This prospective, open-label, single-centre, non-randomised study included glaucoma and healthy subjects. IOP and OPA values acquired by the pressure-measuring contact lens (PMCL) device in one patient's eye at the beginning of the measurement were compared with tonometry values (Goldman applanation tonometry (GAT) and dynamic contour tonometry (DCT)) in the same eye just before PMCL placement. Furthermore, IOP and OPA values measured with PMCL on the study eye during a water drinking test (WDT) were compared with DCT values in the fellow eye. Comparisons were performed using t-tests with 95% Confidence Intervals.

RESULTS

Twenty-four-hour IOP and OPA curves were obtained for eight subjects. The mean IOP difference between PMCL and tonometry on the same eye was within ±5 mm Hg in 75% (GAT) and 87.5% (DCT) of subjects. IOP variations due to WDT were detected by PMCL and DCT, showing an average increase of 2.43 and 1.85 mm Hg, respectively. Differences between PMCL and DCT for IOP variations in fellow eyes were within ±5 mm Hg for 97.2% of time points. The difference between OPA in fellow eyes was within ±5 mm Hg for 85.5% of the time points.

CONCLUSIONS

This first-in-human study is a proof-of-concept for 24-hour continuous measurements of IOP and OPA with the PMCL. This device is non-invasive and has good comparability with standard tonometry.

摘要

背景/目的:本研究评估了一种新型隐形眼镜装置用于 24 小时内连续测量眼压(IOP)和眼脉冲幅度(OPA)的可行性。

方法

本前瞻性、开放标签、单中心、非随机研究纳入了青光眼和健康受试者。在测量开始时,一只眼的压力测量隐形眼镜(PMCL)设备获得的 IOP 和 OPA 值与同一眼在 PMCL 放置前的眼压值(Goldman 压平眼压计(GAT)和动态轮廓眼压计(DCT))进行比较。此外,在饮水试验(WDT)期间,用 PMCL 测量研究眼的 IOP 和 OPA 值与对侧眼的 DCT 值进行比较。比较采用 95%置信区间的 t 检验进行。

结果

8 名受试者获得了 24 小时的 IOP 和 OPA 曲线。在 75%(GAT)和 87.5%(DCT)的受试者中,PMCL 和同眼眼压计之间的平均 IOP 差值在±5mmHg 以内。PMCL 和 DCT 均能检测到 WDT 引起的 IOP 变化,平均分别增加 2.43mmHg 和 1.85mmHg。在对侧眼,PMCL 和 DCT 之间的 IOP 变化差值在 97.2%的时间点在±5mmHg 以内。对侧眼 OPA 的差值在 85.5%的时间点在±5mmHg 以内。

结论

这是首例使用 PMCL 进行 24 小时连续测量 IOP 和 OPA 的人体研究,证明了该设备的概念验证。该设备非侵入性,与标准眼压计具有良好的可比性。