Suppr超能文献

导管肝素化对肺动脉血氧测定的影响。

Effect of heparinization of catheters on pulmonary artery oximetry.

作者信息

Stritter G M, Pearl R G, Mihm F G

机构信息

Department of Anesthesia, Stanford University School of Medicine, CA.

出版信息

J Clin Monit. 1988 Jul;4(3):204-9. doi: 10.1007/BF01621817.

Abstract

A clinical study was performed in two phases to determine whether pulmonary artery oximeter catheters that were impregnated or bonded with heparin would affect the accuracy of measurements of in vivo mixed venous oxygen saturation (Sv-O2). In phase 1, 40 patients were catheterized with either a heparin-impregnated or a plain pulmonary artery catheter. Blood was sampled at random times to correlate in vivo with in vitro Sv-O2 measurements. In phase 2, 16 patients who were not receiving systemic heparin therapy or aspirin and who had no coagulopathies were catheterized with either a heparin-bonded or a plain pulmonary artery catheter in a blinded order. In phase 1, a total of 364 blood samples were obtained from 40 patients. Linear regression analysis of the pooled data demonstrated y = 0.98x - 0.01, r = 0.93, P less than 0.001, and n = 141 with heparin-impregnated catheters; and y = 0.87x + 8.0, r = 0.81, P less than 0.001, and n = 223 with plain catheters. The mean difference (in vivo minus in vitro) revealed a similar error (-1.3 +/- 0.4 versus -1.4 +/- 0.4, respectively, mean +/- SE). The 95% confidence limits of an individual value (+/- 8.1 versus +/- 12.3) suggested slightly greater accuracy with heparin-impregnated catheters. In phase 2, a total of 134 blood samples were obtained from 16 patients. Linear regression analysis showed nearly equal performance with heparin-bonded and plain catheters (r = 0.97 versus r = 0.98, respectively) with similar slopes (1.0 versus 1.1, respectively) but different intercepts (-0.6 versus -8.4, respectively).(ABSTRACT TRUNCATED AT 250 WORDS)

摘要

进行了一项分两个阶段的临床研究,以确定浸渍或结合肝素的肺动脉血氧饱和度导管是否会影响体内混合静脉血氧饱和度(Sv - O2)测量的准确性。在第一阶段,40例患者分别插入浸渍肝素或普通肺动脉导管。在随机时间采集血液样本,以将体内测量结果与体外Sv - O2测量结果进行关联。在第二阶段,16例未接受全身肝素治疗或阿司匹林治疗且无凝血功能障碍的患者,以盲法顺序分别插入结合肝素或普通肺动脉导管。在第一阶段,从40例患者共获取364份血样。对汇总数据进行线性回归分析显示,使用浸渍肝素导管时,y = 0.98x - 0.01,r = 0.93,P < 0.001,n = 141;使用普通导管时,y = 0.87x + 8.0,r = 0.81,P < 0.001,n = 223。平均差异(体内减去体外)显示出相似的误差(分别为-1.3±0.4与-1.4±0.4,均值±标准误)。单个值的95%置信区间(±8.1与±12.3)表明浸渍肝素导管的准确性略高。在第二阶段,从16例患者共获取134份血样。线性回归分析显示,结合肝素导管和普通导管的性能几乎相同(r分别为0.97和0.98),斜率相似(分别为1.0和1.1),但截距不同(分别为-0.6和-8.4)。(摘要截断于250字)

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验