Waller J L, Kaplan J A, Bauman D I, Craver J M
Anesth Analg. 1982 Aug;61(8):676-9.
A new pulmonary arterial catheter (Opticath), containing the standard pulmonary artery catheter features plus two fiberoptic filaments to permit continuous measurement of oxygen saturation (SVO2) by a companion oximeter, was studied in 13 patients undergoing elective coronary bypass surgery. The study was designed to evaluate the accuracy of Opticath SVO2 measurements, to determine the incidence of catheter-related problems, and to correlate changes in SVO2 with hemodynamic changes. A good correlation was found between the SVO2 determined by the Opticath and that measured by American Optical and Radiometer Oximeters (r = 0.92 and 0.89 respectively; p less than 0.0001). There was a significant correlation between increases or decreases in values of SVO2 greater than or equal to 5% and corresponding changes in cardiac index, stroke index, and left ventricular stroke work index (p less than 0.001), and an 86% probability that SVO2 decrease greater than or equal to 5% reflected a significant decline in cardiac index.
一种新型肺动脉导管(Opticath),具备标准肺动脉导管的特性,外加两根光纤丝,可通过配套的血氧计持续测量血氧饱和度(SVO2)。对13例择期冠状动脉搭桥手术患者使用该导管进行了研究。本研究旨在评估Opticath测量SVO2的准确性,确定导管相关问题的发生率,并将SVO2的变化与血流动力学变化相关联。发现Opticath测定的SVO2与美国光学公司和Radiometer血氧计测量的SVO2之间具有良好的相关性(分别为r = 0.92和0.89;p小于0.0001)。SVO2值增加或减少大于或等于5%与心脏指数、每搏指数和左心室每搏功指数的相应变化之间存在显著相关性(p小于0.001),且SVO2下降大于或等于5%有86%的概率反映心脏指数显著下降。