Suppr超能文献

前额脉搏血氧饱和度测定与手指脉搏血氧饱和度测定及动脉血气测量的比较。

Forehead pulse oximetry compared with finger pulse oximetry and arterial blood gas measurement.

作者信息

Cheng E Y, Hopwood M B, Kay J

机构信息

Department of Anesthesiology, Medical College of Wisconsin, Milwaukee 53226.

出版信息

J Clin Monit. 1988 Jul;4(3):223-6. doi: 10.1007/BF01621821.

Abstract

Usual monitoring sites for pulse oximetry involve the fingers, toes, ear lobe, and nasal septum. This study examined the performance of a forehead sensor compared with a finger sensor for the pulse oximeter and arterial blood gas (ABG) analysis. Ten healthy adult volunteers and 22 ventilator-dependent patients were studied. The arterial oxygen saturation detected by forehead pulse oximetry (SpO2) correlated well with finger SpO2 and arterial oxygen saturation (SaO2) determined by arterial blood gas analysis in the healthy volunteers. Forehead SpO2 in mechanically ventilated patients correlated well with finger SpO2 and SaO2 when heart rate detected by pulse oximeter differed less than 10% from apical heart rate. Factors that caused a difference in oximeter-detected heart rate and apical heart rate were extensive tissue edema, head movement, and difficulty securing good tape placement. This suggests that when signal strength is weak, causing poor pulse rate detection, there will also be problems associated with accurate SpO2. The forehead pulse oximeter sensor works well on healthy, well-oxygenated volunteers. Difficulty was experienced when applying and using the sensor on critically ill patients. The reliability of the forehead pulse oximeter sensor has not been established at low saturations.

摘要

脉搏血氧饱和度仪的常用监测部位包括手指、脚趾、耳垂和鼻中隔。本研究比较了前额传感器与手指传感器在脉搏血氧饱和度仪及动脉血气(ABG)分析中的性能。研究对象包括10名健康成年志愿者和22名依赖呼吸机的患者。在前额脉搏血氧饱和度仪检测的动脉血氧饱和度(SpO2)与健康志愿者手指SpO2以及通过动脉血气分析测定的动脉血氧饱和度(SaO2)之间存在良好的相关性。当脉搏血氧饱和度仪检测的心率与心尖心率相差小于10%时,机械通气患者的前额SpO2与手指SpO2和SaO2之间具有良好的相关性。导致血氧饱和度仪检测的心率与心尖心率出现差异的因素包括广泛的组织水肿、头部移动以及难以确保胶带粘贴良好。这表明,当信号强度较弱导致脉搏率检测不佳时,SpO2的准确性也会出现问题。前额脉搏血氧饱和度仪传感器在健康、氧合良好的志愿者身上效果良好。在危重症患者身上应用和使用该传感器时遇到了困难。在前额脉搏血氧饱和度仪传感器在低饱和度情况下的可靠性尚未得到证实。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验