J Clin Monit. 1988 Jul;4(3):223-6. doi: 10.1007/BF01621821.

Abstract

Usual monitoring sites for pulse oximetry involve the fingers, toes, ear lobe, and nasal septum. This study examined the performance of a forehead sensor compared with a finger sensor for the pulse oximeter and arterial blood gas (ABG) analysis. Ten healthy adult volunteers and 22 ventilator-dependent patients were studied. The arterial oxygen saturation detected by forehead pulse oximetry (SpO2) correlated well with finger SpO2 and arterial oxygen saturation (SaO2) determined by arterial blood gas analysis in the healthy volunteers. Forehead SpO2 in mechanically ventilated patients correlated well with finger SpO2 and SaO2 when heart rate detected by pulse oximeter differed less than 10% from apical heart rate. Factors that caused a difference in oximeter-detected heart rate and apical heart rate were extensive tissue edema, head movement, and difficulty securing good tape placement. This suggests that when signal strength is weak, causing poor pulse rate detection, there will also be problems associated with accurate SpO2. The forehead pulse oximeter sensor works well on healthy, well-oxygenated volunteers. Difficulty was experienced when applying and using the sensor on critically ill patients. The reliability of the forehead pulse oximeter sensor has not been established at low saturations.

摘要

脉搏血氧饱和度仪的常用监测部位包括手指、脚趾、耳垂和鼻中隔。本研究比较了前额传感器与手指传感器在脉搏血氧饱和度仪及动脉血气(ABG)分析中的性能。研究对象包括10名健康成年志愿者和22名依赖呼吸机的患者。在前额脉搏血氧饱和度仪检测的动脉血氧饱和度(SpO2)与健康志愿者手指SpO2以及通过动脉血气分析测定的动脉血氧饱和度(SaO2)之间存在良好的相关性。当脉搏血氧饱和度仪检测的心率与心尖心率相差小于10%时,机械通气患者的前额SpO2与手指SpO2和SaO2之间具有良好的相关性。导致血氧饱和度仪检测的心率与心尖心率出现差异的因素包括广泛的组织水肿、头部移动以及难以确保胶带粘贴良好。这表明,当信号强度较弱导致脉搏率检测不佳时,SpO2的准确性也会出现问题。前额脉搏血氧饱和度仪传感器在健康、氧合良好的志愿者身上效果良好。在危重症患者身上应用和使用该传感器时遇到了困难。在前额脉搏血氧饱和度仪传感器在低饱和度情况下的可靠性尚未得到证实。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验