Measurement of crystalline lens tilt in high myopic eyes before cataract surgery using swept-source optical coherence tomography.

作者信息

Lu Qiang, He Wenwen, Qian Dongjin, Lu Yi, Zhu Xiangjia

机构信息

1Department of Ophthalmology, Eye and Ear, Nose, and Throat Hospital, Fudan University, 83 Fenyang Road, Shanghai, 200031 People's Republic of China.

2Eye Institute, Eye and Ear, Nose, and Throat Hospital of Fudan University, Shanghai, 200031 People's Republic of China.

出版信息

Eye Vis (Lond). 2020 Mar 6;7:14. doi: 10.1186/s40662-020-00176-5. eCollection 2020.

Abstract

BACKGROUND

To measure the crystalline lens tilt in eyes with various degrees of myopia before cataract surgery using swept-source optical coherence tomography (SS-OCT).

METHODS

We used SS-OCT (IOLMaster 700) to scan 131 emmetropic eyes (axial length < 24.5 mm), 25 mild/moderate myopic eyes (axial length 24.5-26 mm), and 123 high myopic eyes (52, 29, and 42 eyes with axial lengths of 26-28, 28-30, and > 30 mm, respectively) as part of the routine preoperative examination before cataract surgery. SS-OCT involved B-scans along six meridians. The data were analyzed to assess the magnitude and orientation of the lens tilt and their correlation with other optical biometric parameters.

RESULT

The mean tilt was 3.36 ± 0.98° in emmetropic eyes, 3.07 ± 1.04° in mild/medium myopic eyes, and 2.35 ± 1.01° in high myopic eyes. Tilt correlated significantly and inversely with axial length (Pearson's  = - 0.427,  < 0.001). The crystalline lens tilt predominantly faced the upper outer quadrant relative to the visual axis, symmetrically in both eyes, with mean angles of 24.32° and 147.36° in the right and left eyes, respectively. The variability in the lens tilt direction increased with increasing axial length (χ test,  < 0.001).

CONCLUSION

The magnitude of crystalline lens tilt decreased with increasing axial length. The direction of tilt was predominantly towards the upper outer quadrant in both eyes. The variability in the tilt orientation increased with increasing axial length.

TRIAL REGISTRATION

NIH (clinicaltrial.gov), NCT03062085. Registered 23 February 2017.

摘要

背景

使用扫频光学相干断层扫描(SS-OCT)测量白内障手术前不同近视程度眼睛的晶状体倾斜度。

方法

我们使用SS-OCT(IOLMaster 700)对131只正视眼(眼轴长度<24.5毫米)、25只轻度/中度近视眼(眼轴长度24.5 - 26毫米)和123只高度近视眼(分别为52只、29只和42只眼轴长度为26 - 28毫米、28 - 30毫米和>30毫米的眼睛)进行扫描,作为白内障手术前常规术前检查的一部分。SS-OCT包括沿六个子午线的B扫描。对数据进行分析,以评估晶状体倾斜度的大小和方向及其与其他光学生物测量参数的相关性。

结果

正视眼的平均倾斜度为3.36±0.98°,轻度/中度近视眼为3.07±1.04°,高度近视眼为2.35±1.01°。倾斜度与眼轴长度呈显著负相关(Pearson's = - 0.427,<0.001)。晶状体倾斜主要朝向视轴相对的上外象限,双眼对称,右眼和左眼的平均角度分别为24.32°和147.36°。晶状体倾斜方向的变异性随眼轴长度增加而增加(χ检验,<0.001)。

结论

晶状体倾斜度的大小随眼轴长度增加而减小。倾斜方向主要朝向双眼的上外象限。倾斜方向的变异性随眼轴长度增加而增加。

试验注册

美国国立卫生研究院(clinicaltrial.gov),NCT03062085。2017年2月23日注册。