J Cataract Refract Surg. 2019 Jan;45(1):35-40. doi: 10.1016/j.jcrs.2018.08.025. Epub 2018 Oct 8.
To evaluate crystalline lens and intraocular lens (IOL) tilt using a swept-source optical coherence tomography (SS-OCT) biometer (IOLMaster 700).
Cullen Eye Institute, Baylor College of Medicine, Houston, Texas, USA.
Retrospective case series.
The study patients were evaluated for 1) repeatability of the crystalline lens tilt measurement, 2) preoperative crystalline lens and postoperative IOL tilt in right eyes, 3) mirror symmetry between right eyes and left eyes, 4) the correlation between preoperative crystalline lens and postoperative IOL tilt, and 5) the correlation between the magnitude of lens tilt and ocular parameters.
The study comprised 333 patients. The repeatability was 0.1 degrees for tilt magnitude and 3.0 degrees for tilt direction. Both the crystalline lens and IOL had anterior tilt of the nasal portion with mean tilt magnitudes of 3.7 degrees ± 1.1 (SD) (range 0.4 to 6.9 degrees) for the crystalline lens and 4.9 ± 1.8 degrees (range 1.6 to 10.7 degrees) for the IOL. There was mirror symmetry between right eyes and left eyes. The mean IOL tilt magnitude exceeded crystalline lens tilt by 1.2 ± 1.1 degrees (range -3.2 to 4.0 degrees), and the 2 values were significantly correlated (all P < .01). The magnitude of crystalline lens tilt significantly increased with decreasing axial length and with increasing angle α (all P < .05).
The repeatability of crystalline lens tilt measurements using the SS-OCT biometer was excellent. There was mirror symmetry between the right eyes and left eyes. Preoperative crystalline lens tilt could be used to predict the postoperative IOL tilt. The lens tilt magnitude was greater in short eyes and in eyes with larger angle α.
使用扫频源光学相干断层扫描仪(SS-OCT)生物测量仪(IOLMaster 700)评估晶状体和人工晶状体(IOL)倾斜度。
美国得克萨斯州休斯顿贝勒医学院 Cullen 眼科研究所。
回顾性病例系列。
对研究患者进行了以下评估:1)晶状体倾斜测量的可重复性,2)右眼术前晶状体和术后 IOL 倾斜,3)右眼和左眼之间的镜像对称,4)术前晶状体与术后 IOL 倾斜的相关性,5)晶状体倾斜程度与眼部参数的相关性。
该研究共纳入 333 例患者。倾斜幅度的重复性为 0.1 度,倾斜方向的重复性为 3.0 度。晶状体和 IOL 的鼻侧均有前向倾斜,晶状体倾斜幅度的平均值为 3.7°±1.1(SD)(范围 0.4 至 6.9°),IOL 的倾斜幅度平均值为 4.9°±1.8 度(范围 1.6 至 10.7 度)。右眼和左眼之间存在镜像对称。IOL 倾斜幅度的平均值比晶状体倾斜幅度大 1.2°±1.1 度(范围 -3.2 至 4.0 度),且这 2 个值显著相关(均 P<.01)。晶状体倾斜幅度随眼轴长度的缩短和角度α的增大而显著增加(均 P<.05)。
使用 SS-OCT 生物测量仪测量晶状体倾斜的重复性非常好。右眼和左眼之间存在镜像对称。术前晶状体倾斜可用于预测术后 IOL 倾斜。短眼和角度α较大的眼晶状体倾斜幅度更大。