在改良快速序列诱导期间进行神经肌肉监测:TOF-Cuff® 和 TOF-Scan® 的比较。

Neuromuscular monitoring during modified rapid sequence induction: A comparison of TOF-Cuff® and TOF-Scan®.

机构信息

Institute for Anesthesia and Intensive Care Medicine, Spital Thurgau Frauenfeld, Frauenfeld, Switzerland.

Institute for Anesthesia and Intensive Care Medicine, Spital Thurgau Frauenfeld, Frauenfeld, Switzerland.

出版信息

Australas Emerg Care. 2020 Dec;23(4):217-220. doi: 10.1016/j.auec.2020.02.005. Epub 2020 Mar 12.

Abstract

BACKGROUND

Acceleromyometry is the clinical standard for quantitative neuromuscular monitoring, mostly using the stimulation pattern train-of-four (TOF). TOF-Cuff®, a recently introduced neuromuscular monitor with stimulating electrodes integrated within a blood pressure cuff, assesses the muscular response in the upper arm.

METHODS

The time from administration of a neuromuscular blocking agent to TOF-ratio 0% during modified rapid sequence induction was compared between TOF-Cuff® and acceleromyometry (TOF-Scan®). Included were 26 adults with body mass index <35 kg/m. TOF-Scan® and TOF-Cuff® were simultaneously fitted on patients' opposite arms. The mean difference to TOF-ratio 0% was compared using the one sample t-test (p < 0.05) and Bland-Altman plots.

RESULTS

After anesthesia induction, atracurium 0.9 mg/kg (±0.08) i.v. was administered. The mean time to TOF ratio 0% for TOF-Scan® was 140.4 s (±34.3), and 132.7 s (±32.5) for TOF-Cuff®, with a mean difference of 5.4 (95% CI: -9.9 to 20.7, p = 0.472). The maximum difference between the two modalities was 135 s when the TOF-Cuff® was faster and 60 s when the TOF-Scan® was faster.

CONCLUSIONS

No statistically significant systematic difference was found between TOF-Scan® and TOF-Cuff®. However, there was high variability and wide limits of agreement. The two devices cannot be used interchangeably.

摘要

背景

加速肌电图是定量神经肌肉监测的临床标准,主要使用刺激模式四脉冲串(TOF)。TOF-Cuff®是一种最近引入的神经肌肉监测器,其刺激电极集成在血压袖带内,评估上臂的肌肉反应。

方法

在改良快速序列诱导过程中,从给予神经肌肉阻滞剂到 TOF 比 0%的时间,TOF-Cuff®和加速肌电图(TOF-Scan®)之间进行了比较。纳入 26 名 BMI<35 kg/m 的成年人。TOF-Scan®和 TOF-Cuff®同时安装在患者的对侧手臂上。使用单样本 t 检验(p<0.05)和 Bland-Altman 图比较 TOF 比值 0%的平均差值。

结果

麻醉诱导后,静脉注射 0.9 mg/kg(±0.08)阿曲库铵。TOF-Scan®的 TOF 比值 0%的平均时间为 140.4 s(±34.3),TOF-Cuff®为 132.7 s(±32.5),平均差异为 5.4(95%CI:-9.9 至 20.7,p=0.472)。两种方式之间的最大差异为 135 s,当 TOF-Cuff®更快时,为 60 s,当 TOF-Scan®更快时。

结论

TOF-Scan®和 TOF-Cuff®之间未发现统计学上显著的系统差异。然而,存在高度的可变性和广泛的一致性界限。这两种设备不能互换使用。