Department of Radiotherapy and Radiooncology, University Medical Center Giessen-Marburg, Marburg, Germany.
Institute of Medical Physics and Radiation Protection, University of Applied Sciences, Giessen, Germany.
J Appl Clin Med Phys. 2020 Mar;21(3):52-61. doi: 10.1002/acm2.12824.
In radiation therapy, a Computed Tomography (CT) image is needed for an accurate dose calculation. To allow such a calculation, the CT image values have to be converted into relative electron densities. Thus, standard procedure is to calibrate the CT numbers to relative electron density (RED) by using a phantom with known composition inserts. This calibration curve is energy and CT dependent, therefore most radiotherapy CT acquisitions are obtained with 120 kVp, as each tube voltage needs an additional calibration curve. The commercially available DirectDensity (DD) reconstruction algorithm presents a reconstruction implementation without any dependence on the tube voltage. In comparison, it allows a calibration curve that is directly proportional to the RED, reducing the need of more than one calibration curve. This could potentially optimize CT acquisitions and reducing the dose given to the patient. Three different phantoms were used to evaluate the DirectDensity algorithm in simple and anthropomorphic geometries, as well as setups with metal implants. Scans with the DD algorithm were performed for 80, 100, 120, and 140 kVp. As reference a scan with the standard 120 kVp scan was used. Radiotherapy photon plans were optimized and calculated on the reference image and then transferred to the DD images, where they were recalculated. The dose distributions obtained this way were compared to the reference dose. Differences were found mainly in pure air and high density materials such as bones. The difference of the mean dose was below 0.7%, in most cases below 0.4%. No indication was found that the algorithm is corrupted by metal inserts, enabling the application for all clinical cases. This algorithm offers more variability in CT parameters for radiation therapy and thus a more personalized image acquisition with a high image quality and a lower dose exposure at a robust clinical workflow.
在放射治疗中,需要进行计算机断层扫描(CT)图像以进行准确的剂量计算。为了进行这种计算,必须将 CT 图像的值转换为相对电子密度。因此,标准程序是使用具有已知组成插入物的体模来校准 CT 数与相对电子密度(RED)。该校准曲线取决于能量和 CT,因此大多数放射治疗 CT 采集都是使用 120 kVp 进行的,因为每种管电压都需要额外的校准曲线。市售的 DirectDensity(DD)重建算法提供了一种与管电压无关的重建实现。相比之下,它允许校准曲线与 RED 直接成正比,从而减少了对多个校准曲线的需求。这有可能优化 CT 采集并减少患者接受的剂量。使用三种不同的体模来评估简单和拟人化几何形状以及带有金属植入物的设置中的 DirectDensity 算法。使用 DD 算法对 80、100、120 和 140 kVp 进行了扫描。作为参考,使用了标准的 120 kVp 扫描。对参考图像进行了放射治疗光子计划的优化和计算,然后将其转移到 DD 图像,在那里重新进行了计算。以这种方式获得的剂量分布与参考剂量进行了比较。差异主要出现在纯空气和高密度材料(如骨骼)中。在大多数情况下,平均剂量的差异低于 0.4%。没有发现该算法被金属插入物损坏的迹象,从而可以将其应用于所有临床病例。该算法为放射治疗提供了更多的 CT 参数变化,从而可以进行更高质量的个性化图像采集,并在稳健的临床工作流程中降低剂量暴露。