Central corneal thickness influences intraocular pressure (IOP) measurement. We examined the central corneal thickness of subjects in the Ocular Hypertension Treatment Study (OHTS) and determined if central corneal thickness is related to race.
Cross-sectional study.
One thousand three hundred one OHTS subjects with central corneal thickness measurements.
Central corneal thickness was determined with ultrasonic pachymeters of the same make and model at all clinical sites of the OHTS.
Correlation of mean central corneal thickness with race, baseline IOP, refraction, age, gender, systemic hypertension, and diabetes.
Mean central corneal thickness was 573.0 ± 39.0 μm. Twenty-four percent of the OHTS subjects had central corneal thickness > 600 μm. Mean central corneal thickness for African American subjects (555.7 ± 40.0 μm; n = 318) was 23 μm thinner than for white subjects (579.0 ± 37.0 μm; P < 0.0001). Other factors associated with greater mean central corneal thickness were younger age, female gender, and diabetes.
OHTS subjects have thicker corneas than the general population. African American subjects have thinner corneas than white subjects in the study. The effect of central corneal thickness may influence the accuracy of applanation tonometry in the diagnosis, screening, and management of patients with glaucoma and ocular hypertension.
中央角膜厚度会影响眼内压(IOP)的测量。我们检查了眼高压治疗研究(OHTS)中受试者的中央角膜厚度,并确定中央角膜厚度是否与种族有关。
横断面研究。
1301 名 OHTS 受试者,均进行中央角膜厚度测量。
所有 OHTS 临床站点均采用相同制造商和型号的超声角膜测厚仪测量中央角膜厚度。
平均中央角膜厚度与种族、基线 IOP、屈光度、年龄、性别、系统性高血压和糖尿病的相关性。
平均中央角膜厚度为 573.0 ± 39.0 μm。24%的 OHTS 受试者中央角膜厚度>600 μm。非裔美国受试者(555.7 ± 40.0 μm;n=318)的平均中央角膜厚度比白人受试者(579.0 ± 37.0 μm;P<0.0001)薄 23 μm。其他与平均中央角膜厚度较大相关的因素包括年龄较小、女性和糖尿病。
OHTS 受试者的角膜比一般人群厚。在这项研究中,非裔美国受试者的角膜比白人受试者薄。中央角膜厚度的影响可能会影响青光眼和眼高压患者诊断、筛查和管理中压平眼压计的准确性。