University Eye Clinic, IRCCS, Policlinico San Matteo, Pavia, Italy.
Eye Clinic, Melegnano Hospital, Via Pandina, 1, 20077, Vizzolo Predabissi, Italy.
Graefes Arch Clin Exp Ophthalmol. 2024 Jun;262(6):1857-1863. doi: 10.1007/s00417-023-06354-0. Epub 2024 Jan 11.
To compare central corneal thickness (CCT) measured with optical coherence tomography (OCT) using an automatic algorithm (A-OCT) vs. manual measurements (M-OCT) with respect to the gold standard ultrasound pachymetry (USP).
CCT measurements were performed on both eyes of 28 healthy subjects at four times of the day. A-OCT used an automatic software analysis of the corneal image, M-OCT was performed by two operators by setting a digital calliper on the corneal borders, and USP was performed after corneal anesthesia. Measurements were compared using a three-way repeated measures ANOVA. Bland-Altman plots were used to evaluate the agreement between OCT measurements and USP.
Both A-OCT and M-OCT significantly underestimated the USP measures, with the mean difference, i.e., the systematic error, being larger for A-OCT (- 19.0 µm) than for M-OCT (- 6.5 µm). Good reproducibility between the two operators was observed. Bland-Altman plots showed that both OCT methods suffered from proportional errors, which were not affected by time and eye.
Measuring CCT with OCT yielded lower values than USP. Therefore, clinicians should be aware that corneal thickness values may be influenced by the measurement method and that the various devices should not be used interchangeably in following up a given patient. Intriguingly, M-OCT had less systematic error than A-OCT, an important outcome that clinicians should consider when deciding to use an OCT device.
比较使用自动算法(A-OCT)和手动测量(M-OCT)光学相干断层扫描(OCT)测量中央角膜厚度(CCT)与超声角膜测厚术(USP)金标准的差异。
对 28 名健康受试者的双眼在一天中的四个时间点进行 CCT 测量。A-OCT 使用角膜图像的自动软件分析,M-OCT 由两名操作员通过在角膜边界设置数字卡尺进行测量,USP 在角膜麻醉后进行。采用三向重复测量方差分析比较测量值。Bland-Altman 图用于评估 OCT 测量值与 USP 的一致性。
A-OCT 和 M-OCT 均显著低估了 USP 测量值,其中 A-OCT(-19.0μm)的平均差异,即系统误差,大于 M-OCT(-6.5μm)。两名操作员之间具有良好的可重复性。Bland-Altman 图显示两种 OCT 方法均存在比例误差,不受时间和眼睛的影响。
OCT 测量 CCT 的值低于 USP。因此,临床医生应注意角膜厚度值可能受到测量方法的影响,并且在随访特定患者时,不应将各种设备互换使用。有趣的是,M-OCT 的系统误差小于 A-OCT,这是临床医生在决定使用 OCT 设备时应考虑的重要结果。