Br J Ophthalmol. 2021 Jan;105(1):22-26. doi: 10.1136/bjophthalmol-2019-314954. Epub 2020 Mar 26.
To assess the repeatability and reliability of the most important tomographic parameters for characterising keratoconus measured with a Pentacam HR (high resolution).
Overall, 230 eyes in 158 patients with keratoconus were analysed. We performed five consecutive corneal tomography examinations for each eye with a Pentacam HR in patients with keratoconus. Study eyes were classified into three groups depending on the maximum posterior elevation (max_BFS_post): grade 1 for cases of keratoconus with a max_BFS_post of 40 µm; grade 2 for those with a max_BFS_post of between 41 and 75 µm and grade 3 for those with a max_BFS_post of over 75 µm. We calculated the intraclass correlation coefficients (ICCs) and repeatability limits of parameters from tomography and aberrometry.
All the parameters were found to have excellent ICCs (0.9). The repeatability limits for the key parameters were higher than 0.5D for the power parameters, 20° for the axis of corneal astigmatism and 10 µm for the thinnest corneal thickness. Further, we obtained repeatability limits of above 0.1 µm for the aberrometry values and overall greater than 15° for the coma axis. All the values increase with the severity of keratoconus, except for that of the coma axis which falls with keratoconus grade.
The reliability indicated by ICCs supports the view that the Pentacam HR is useful for the diagnosis of keratoconus. The repeatability limits suggest that new criteria should be established for monitoring progression taking into account the real measurements that can be made using this system.
评估 Pentacam HR(高分辨率)测量圆锥角膜的最重要的层析参数的重复性和可靠性。
共分析了 158 例 230 只圆锥角膜患者的眼睛。对圆锥角膜患者的每只眼进行了 5 次连续的 Pentacam HR 角膜层析检查。研究眼根据最大后表面隆起(max_BFS_post)分为 3 组:max_BFS_post 为 40µm 的圆锥角膜为 1 级;max_BFS_post 为 41-75µm 的为 2 级;max_BFS_post 超过 75µm 的为 3 级。我们计算了层析和像差参数的组内相关系数(ICCs)和重复性限。
所有参数的 ICCs 均为优秀(0.9)。关键参数的重复性限为:屈光力参数高于 0.5D,角膜散光轴位高于 20°,最薄角膜厚度高于 10µm。此外,我们还获得了像差值的重复性限高于 0.1µm,彗差轴位总体上大于 15°。所有值都随着圆锥角膜的严重程度增加而增加,除了彗差轴位随着圆锥角膜等级的增加而下降。
由 ICCs 表示的可靠性支持了 Pentacam HR 对圆锥角膜诊断有用的观点。重复性限表明,应根据该系统可进行的实际测量值,为监测进展制定新的标准。