Evaluation of the Reliability and Repeatability of Scheimpflug System Measurement in Keratoconus.

作者信息

de Luis Eguileor Beatriz, Escudero Argaluza Julene, Pijoán Zubizarreta Jose Ignacio, Santamaria Carro Alaitz, Etxebarria Ecenarro Jaime

机构信息

Hospital Universitario Cruces, Servicio de Oftalmología, Barakaldo, Spain.

Unidad de Epidemiología Clínica, Hospital Universitario Cruces/Biocruces Health Research Institute, Barakaldo, Spain.

出版信息

Cornea. 2018 Feb;37(2):177-181. doi: 10.1097/ICO.0000000000001373.

Abstract

PURPOSE

To assess the reliability and repeatability of the parameters obtained using a Scheimpflug system (Pentacam HR-high resolution) for monitoring progression in eyes with keratoconus.

METHODS

A prospective study was conducted based on measurements taken with the Pentacam HR in 82 eyes with keratoconus. We calculated the intraclass correlation coefficient (ICC) and the repeatability limit for topographic and aberrometric parameters.

RESULTS

All the parameters analyzed had excellent ICCs, above 0.95 in all cases, except for the measurement of the anterior best-fit sphere, for which the ICC was 0.873. The key repeatability limits for the parameters analyzed were as follows: 1.54 diopters (D) for the maximum keratometry in the central cornea, 1.08 D for maximum keratometry, 32.6 degrees for the axis of corneal astigmatism, 11.9 degrees for the coma axis, and 14.2 μm for thinnest corneal thickness.

CONCLUSIONS

The level of repeatability of the measurements obtained with a Pentacam HR confirms the usefulness of this system for the diagnosis of keratoconus. Nevertheless, it has certain limitations for monitoring individual progression of keratoconus and planning intrastromal corneal ring segment insertion.

摘要

目的

评估使用眼前节分析系统(Pentacam HR - 高分辨率)获得的参数在监测圆锥角膜患者病情进展方面的可靠性和可重复性。

方法

对82只圆锥角膜患眼进行前瞻性研究,使用Pentacam HR进行测量。我们计算了地形学和像差测量参数的组内相关系数(ICC)和可重复性极限。

结果

除前表面最佳拟合球镜度测量的ICC为0.873外,所有分析参数的ICC均极佳,在所有情况下均高于0.95。所分析参数的关键可重复性极限如下:中央角膜最大角膜曲率为1.54屈光度(D),最大角膜曲率为1.08 D,角膜散光轴为32.6度,彗差轴为11.9度,最薄角膜厚度为14.2μm。

结论

Pentacam HR获得的测量结果的可重复性水平证实了该系统在圆锥角膜诊断中的实用性。然而,在监测圆锥角膜个体病情进展和规划基质内环植入方面存在一定局限性。