To evaluate the repeatability and agreement of two different ocular biometers and Scheimpflug devices in keratoconus eyes.
This prospective, comparative trial took place at the University hospital, Goethe University, Frankfurt am Main, Germany. We included eyes with keratoconus, one eye per patient, randomly selected. Measurements were taken with Galilei G6 (Ziemer, Switzerland) and Pentacam AXL (Oculus, Germany), three consecutive measurements each. Repeatability and agreement were evaluated for simulated keratometry (simK), astigmatism (simA), maximum keratometry (KMax) and its axis, total keratometry (TCP), axial length (AL), anterior chamber depth (ACD), and thinnest pachymetry (TCT).
Both devices showed an excellent repeatability with intra class correlation (ICC) of > 0.97 for all parameters. The 95% limits of repeatability (LoR95%) and agreement (LoA95%) were narrow for all parameters. The Galilei G6 had a narrower LoAR95% for TCT (2.1 μm vs. 4.6 μm), but a wider LoR95% for KMax (0.52D vs. 1.18D). No relevant difference was found for the other parameters. Agreement between the devices was good to moderate, especially for simK and TCP.
Both devices show excellent repeatability with narrow LoR95% and high ICC for all parameters. The only relevant difference was found for KMax and TCT in favor of Pentacam AXL and Galilei G6, respectively. Agreement was good to moderate, and most parameters should not be considered interchangeable.
评估两种不同的眼生物测量仪和 Scheimpflug 设备在圆锥角膜眼中的重复性和一致性。
本前瞻性、对照试验在德国法兰克福歌德大学附属医院进行。我们纳入了每例患者的一只圆锥角膜眼,随机选择。使用 Galilei G6(瑞士 Ziemer)和 Pentacam AXL(德国 Oculus)分别进行三次连续测量。评估模拟角膜曲率(simK)、散光(simA)、最大角膜曲率(KMax)及其轴、总角膜曲率(TCP)、眼轴(AL)、前房深度(ACD)和最薄角膜厚度(TCT)的重复性和一致性。
两种设备的所有参数的组内相关系数(ICC)均>0.97,显示出极好的重复性。所有参数的 95%重复性限(LoR95%)和一致性限(LoA95%)均较窄。Galilei G6 对于 TCT 的 LoAR95%较窄(2.1μm 比 4.6μm),但对于 KMax 的 LoR95%较宽(0.52D 比 1.18D)。其他参数未发现明显差异。两种设备之间的一致性为良好到中度,尤其是对于 simK 和 TCP。
两种设备的所有参数均具有极好的重复性,LoR95%较窄,ICC 较高。仅在 KMax 和 TCT 方面发现了明显差异,Pentacam AXL 优于 Galilei G6。一致性为良好到中度,大多数参数不应视为可互换。