Fritsch R, Hausmann R
Chirurgischen Veterinärklinik und Poliklinik, Justus-Liebig-Universität Giessen.
Tierarztl Prax. 1988;16(4):373-6.
In 26 horses measurings of blood pressure, using the Dinamap 1255 research monitor, were performed and compared for accuracy with blood pressure values obtained by direct measurement in the arteria carotis. It was found that oscillometrically ascertained initial blood values in standing horses have a sufficient correlation to initial blood pressure values obtained by direct measurement. But every blood pressure change is registered too late, when using the oscillometric blood pressure apparatus. This temporal delay becomes unacceptable, when a considerable increase in blood pressure occurs. A reliable oscillometric blood pressure measurement with uneven and irregular pulse waves is not possible at all.
使用Dinamap 1255研究监测仪对26匹马进行了血压测量,并将测量结果与通过直接测量颈动脉获得的血压值进行了准确性比较。结果发现,站立马匹通过示波法确定的初始血压值与通过直接测量获得的初始血压值具有足够的相关性。但是,使用示波血压仪时,每次血压变化的记录都太晚了。当血压大幅升高时,这种时间延迟变得不可接受。对于不均匀且不规则的脉搏波,根本无法进行可靠的示波血压测量。