Dinamap监护仪测量婴幼儿血压的准确性。

Accuracy of blood pressure measurement by the Dinamap monitor in infants and children.

作者信息

Park M K, Menard S M

出版信息

Pediatrics. 1987 Jun;79(6):907-14.

PMID:3588145
Abstract

Indirect BPs measured by the Dinamap monitor, an oscillometric device, and the conventional auscultatory method were compared with the direct radial artery pressure in infants and children. There were 29 patients in the Dinamap group with a median age of 18 months (ranges 1 month to 16 years) and 20 patients in the auscultatory group with a median age of 3.5 years (ranges 3 months to 16 years). The direct radial artery pressures were recorded on a strip chart and the ranges of pressures were obtained for systolic, diastolic, and mean pressures. The range of the direct readings was converted to a weighted single reading, and comparisons were made between the direct and indirect readings. The linear regression equation between the direct (x) and the Dinamap (y) readings (with correlation coefficient) was y = 1.05x - 5.36 (r = .97) for systolic, y = 1.10x - 4.65 (r = .903) for diastolic, and y = 1.06x - 4.21 (r = .917) for mean pressures. The linear regression equation (with correlation coefficient) for the auscultatory pressure was y = 1.60x - 68.23 (r = .872) and y = 1.38x - 16.47 (r = .874) for diastolic pressures. The error was defined as the indirect reading minus the direct reading. The mean error +/- SD was -0.24 +/- 3.26 for systolic, 1.28 +/- 4.74 for diastolic, and 0.10 +/- 4.56 mm Hg for mean pressures by the Dinamap method. It was -1.65 +/- 6.68 for systolic, and 8.70 +/- 5.97 for diastolic pressures by the auscultatory method.(ABSTRACT TRUNCATED AT 250 WORDS)

摘要

将通过振荡式设备Dinamap监护仪测量的间接血压以及传统听诊法测量的间接血压,与婴幼儿和儿童的桡动脉直接血压进行比较。Dinamap组有29例患者,中位年龄为18个月(范围1个月至16岁);听诊组有20例患者,中位年龄为3.5岁(范围3个月至16岁)。桡动脉直接血压记录在带状图表上,并获取收缩压、舒张压和平均压的压力范围。将直接读数范围转换为加权单一读数,并对直接读数和间接读数进行比较。直接读数(x)与Dinamap读数(y)之间的线性回归方程(及相关系数)为:收缩压y = 1.05x - 5.36(r = 0.97),舒张压y = 1.10x - 4.65(r = 0.903),平均压y = 1.06x - 4.21(r = 0.917)。听诊法测量血压的线性回归方程(及相关系数)为:舒张压y = 1.60x - 68.23(r = 0.872),舒张压y = 1.38x - 16.47(r = 0.874)。误差定义为间接读数减去直接读数。Dinamap法测量收缩压的平均误差±标准差为-0.24±3.26,舒张压为1.28±4.74,平均压为0.10±4.56 mmHg。听诊法测量收缩压的平均误差为-1.65±6.68,舒张压为8.70±5.9

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验