Suppr超能文献

通过测量种植体位移评估不同骨类型中圆柱形和锥形种植体的初期稳定性:一项研究。

Evaluation of Primary Stability of Cylindrical and Tapered Implants in Different Bone Types by Measuring Implant Displacement: An Study.

作者信息

Sugiura Tsutomu, Yamamoto Kazuhiko, Horita Satoshi, Murakami Kazuhiro, Kirita Tadaaki

机构信息

Department of Oral and Maxillofacial Surgery, Nara Medical University, Kashihara, Nara, Japan.

出版信息

Contemp Clin Dent. 2019 Jul-Sep;10(3):471-476. doi: 10.4103/ccd.ccd_788_18.

Abstract

AIMS

The purposes of this study were to investigate the primary stability of cylindrical and tapered implants in different bone types by measuring implant displacement and to examine the relationship between insertion torque value (ITV) and implant displacement.

MATERIALS AND METHODS

Four different polyurethane bone models consisted of low-density or low-to-medium-density cancellous bone with or without a cortical bone layer. A total of 120 cylindrical and tapered implants were placed into bone blocks ( = 15 in each group), and the ITV was measured. A lateral load of 15 N was applied to the top of the abutment, and implant displacement was recorded.

RESULTS

Implant displacement was significantly affected by cancellous bone density and to a lesser degree by cortical bone thickness. The displacement of tapered implants was significantly smaller than that of cylindrical implants in the presence of cortical bone. However, both implant groups showed similar ITV in the low-density cancellous bone model with the cortical bone layer. There was a correlation between ITV and displacement in the cylindrical and tapered implants. However, no correlation was observed between ITV and displacement within each bone type.

CONCLUSIONS

Implant stability depended mainly on the bone type, whereas implant design had a limited influence on primary stability. The use of tapered implants may be advantageous for improving primary stability in patients with low-density cancellous bone only when crestal cortical bone exists. The same ITV of cylindrical and tapered implants did not necessarily represent similar primary stability in the bone type.

摘要

目的

本研究的目的是通过测量种植体位移来研究圆柱形和锥形种植体在不同骨类型中的初始稳定性,并探讨植入扭矩值(ITV)与种植体位移之间的关系。

材料与方法

四种不同的聚氨酯骨模型,由低密度或低至中密度的松质骨组成,有或没有皮质骨层。将总共120个圆柱形和锥形种植体植入骨块中(每组 = 15个),并测量ITV。在基台顶部施加15 N的侧向载荷,并记录种植体位移。

结果

种植体位移受松质骨密度的显著影响,受皮质骨厚度的影响较小。在存在皮质骨的情况下,锥形种植体的位移明显小于圆柱形种植体。然而,在具有皮质骨层的低密度松质骨模型中,两个种植体组的ITV相似。圆柱形和锥形种植体的ITV与位移之间存在相关性。然而,在每种骨类型中,未观察到ITV与位移之间的相关性。

结论

种植体稳定性主要取决于骨类型,而种植体设计对初始稳定性的影响有限。仅当存在嵴顶皮质骨时,使用锥形种植体可能有利于改善低密度松质骨患者的初始稳定性。圆柱形和锥形种植体相同的ITV在骨类型中不一定代表相似的初始稳定性。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/df67/7150554/a4dce27c8541/CCD-10-471-g001.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验