用于公共卫生站的自动可调血压系统的验证

Validation of an Automated and Adjustable Blood Pressure System for Use with a Public Health Station.

作者信息

Macaulay Timothy R, Erceg David N, McMillan Bjorn, Ramirez Joel E, Dominguez Jesus F, Vallejo Alberto F, Schroeder E Todd

机构信息

Division of Biokinesiology and Physical Therapy, University of Southern California, Los Angeles, CA, USA.

Stayhealthy Inc., Monrovia, CA, USA.

出版信息

Vasc Health Risk Manag. 2020 Apr 9;16:133-142. doi: 10.2147/VHRM.S246401. eCollection 2020.

Abstract

OBJECTIVE

A new automated and adjustable blood pressure (BP) system has been developed to improve the accuracy of BP measurements on public-use health stations. This self-fitting BP system includes a mechanical cuff that wraps down to the user's arm prior to bladder inflation. The purpose of this study was to validate the adaptable BP system (ABPS) using the current standards from the Association for the Advancement of Medical Instrumentation (AAMI).

METHODS

The AAMI/ISO 81060:2013 standards for clinical validation of non-invasive automated arterial BP measurement devices were followed precisely using the same arm sequential method. For each participant, BP was measured over multiple trials by trained observers alternating a reference sphygmomanometer with the ABPS. All study requirements were met with 85 qualifying participants, each with 3 valid paired determinations.

RESULTS

The mean difference between ABPS BP and reference BP using all 255 paired determinations was -2.4 ± 7.7 mmHg for systolic and 1.7 ± 5.7 mmHg for diastolic. The standard deviation of the averaged paired determinations per participant was 6.3 mmHg for systolic and 5.2 mmHg for diastolic. Arm circumference measurements had a mean error of -2.1 ± 2.4 cm (R = 0.87). A new prediction model for arm circumference was validated using a holdout dataset (R = 0.94).

CONCLUSION

The results of the study confirm that the ABPS met all benchmarks established by the AAMI. The device accurately measures BP across a wide range of arm circumferences (24-44 cm) and is suitable for use by individuals to self-monitor BP.

摘要

目的

已开发出一种新型自动可调血压系统,以提高公共卫生站血压测量的准确性。这种自动适配血压系统包括一个机械袖带,在气囊充气前会向下包裹到使用者的手臂上。本研究的目的是根据医学仪器促进协会(AAMI)的现行标准,对这种适配血压系统(ABPS)进行验证。

方法

严格遵循AAMI/ISO 81060:2013标准中关于无创自动动脉血压测量设备临床验证的要求,采用相同手臂顺序测量法。对于每位参与者,由经过培训的观察者使用参考血压计和ABPS交替进行多次测量。85名符合条件的参与者满足了所有研究要求,每位参与者均有3次有效的配对测定。

结果

使用所有255对配对测定数据,ABPS测量的收缩压与参考收缩压的平均差值为-2.4±7.7 mmHg,舒张压的平均差值为1.7±5.7 mmHg。每位参与者平均配对测定的标准差,收缩压为6.3 mmHg,舒张压为5.2 mmHg。臂围测量的平均误差为-2.1±2.4 cm(R = 0.87)。使用一个保留数据集验证了一种新的臂围预测模型(R = 0.94)。

结论

研究结果证实ABPS符合AAMI设定的所有基准。该设备能在较宽的臂围范围(24 - 44 cm)内准确测量血压,适合个人用于自我监测血压。