采用碳纤维增强聚醚醚酮(CFR-PEEK)作为髋臼内衬的全髋关节置换术的令人失望的长期结果:一项平均随访 14.3 年的前瞻性研究。

Disappointing long-term outcome of THA with carbon-fiber-reinforced poly-ether-ether-ketone (CFR-PEEK) as acetabular insert liner: a prospective study with a mean follow-up of 14.3 years.

机构信息

Department of Orthopaedic Surgery, Zuyderland Medical Centre Sittard, The Netherlands.

出版信息

Hip Int. 2021 Nov;31(6):735-742. doi: 10.1177/1120700020918157. Epub 2020 Apr 27.

Abstract

INTRODUCTION

Insert liner wear of the acetabular component is one of the predictive values for survival of total hip arthroplasties (THAs). This prospective single-centre study was designed to evaluate the follow-up of carbon-fiber-reinforced poly-ether-ether-ketone (CFR-PEEK) insert liner used as bearing in cementless THAs.

METHODS

29 healthy patients with an indication for cementless THA were selected for a CFR-PEEK insert liner and followed over time. All patients received a cementless THA with a CFR-PEEK insert liner used as bearing. At different follow-up moments patients were routinely examined and were analysed using the Oxford Hip Score (OHS), the modified Merle d'Aubigne-Postel (MAP) score, and radiologically. At the follow up moments the plain radiographics where assessed for loosening, cyst formations and wear of the CFR-PEEK liners.

RESULTS

At a mean of 14.3 years follow-up 4 revisions of the acetabular component were performed, resulting in a survival rate of 86.5% (CI 95%, 72.4-96.6). A statistically significant difference in OHS and MAP scores between pre- and postoperative follow-up moments was observed. The acetabular components of the remaining patients showed no radiological abnormalities at 14.3 years follow-up. The overall CFR-PEEK wear was low, with a mean of 0.81 (0.2-1.4) mm wear at 14.3 years follow-up.

CONCLUSIONS

In this series we found an aseptic loosening with unclear reasons in 4 well-positioned acetabular components, hence we do not recommend routine use of CFR-PEEK insert liners as bearing in cementless THAs. All the remaining THAs and acetabular components were without abnormalities at 14.3 years follow-up.

摘要

简介

髋臼部件的衬垫磨损是全髋关节置换术 (THA) 生存的预测指标之一。本前瞻性单中心研究旨在评估作为非骨水泥 THA 中轴承使用的碳纤维增强聚醚醚酮 (CFR-PEEK) 衬垫的随访情况。

方法

选择 29 例有非骨水泥 THA 适应证的健康患者使用 CFR-PEEK 衬垫,并随时间进行随访。所有患者均接受了非骨水泥 THA,使用 CFR-PEEK 衬垫作为轴承。在不同的随访时间,患者通常接受检查,并使用牛津髋关节评分 (OHS)、改良的 Merle d'Aubigne-Postel (MAP) 评分和影像学进行分析。在随访时,对 CFR-PEEK 衬垫的松动、囊肿形成和磨损进行平片评估。

结果

平均随访 14.3 年后,进行了 4 次髋臼部件翻修,生存率为 86.5%(CI 95%,72.4-96.6)。OHS 和 MAP 评分在术前和术后随访时间之间存在统计学显著差异。其余患者的髋臼部件在 14.3 年随访时未发现放射学异常。整体 CFR-PEEK 磨损较低,平均为 0.81(0.2-1.4)mm,14.3 年随访时。

结论

在本系列中,我们发现 4 个位置良好的髋臼部件出现无菌性松动,原因不明,因此我们不建议常规在非骨水泥 THA 中使用 CFR-PEEK 衬垫作为轴承。所有其余的 THA 和髋臼部件在 14.3 年随访时均无异常。