Difference in the nocturnal behaviour of blood pressure between monitorings with the arm-cuff method and with the finger-volume oscillometric method.

作者信息

Imai Y, Sasaki S, Minami N, Munakata M, Sekino H, Abe K, Yoshinaga K

机构信息

Department of Medicine, Tohoku University School of Medicine, Sendai, Japan.

出版信息

J Hypertens Suppl. 1988 Dec;6(4):S61-3. doi: 10.1097/00004872-198812040-00015.

Abstract

The nocturnal behaviour of blood pressure obtained by a new portable device for monitoring ambulatory blood pressure (ABPM 630) was compared with that obtained by a finger-volume oscillometric device (BP-100); the former uses a conventional arm-cuff inflated by CO2 gas to eliminate the noise of the motorized pump, and is based on a cuff-oscillometric as well as the Korotkoff sound technique (microphone method). With the microphone method in ABPM 630, the mean differences from the conventional auscultatory method were -0.28 +/- 6.15 mmHg (mean +/- s.d.) for systolic and 0.96 +/- 6.28 mmHg for diastolic blood pressure, and there was a highly significant correlation between blood pressure values measured by the ABPM 630 and by the auscultatory method. In 40 patients with mild to moderate hypertension 24-h blood pressure was monitored simultaneously with the ABPM 630 and BP-100. The daytime average of systolic blood pressure measured with the ABPM 630 was similar to that measured with the BP-100, whereas the night-time average assessed by the former was significantly higher than that assessed by the latter. Arm-cuff inflation by ABPM 630 caused some degree of sleep disturbance in 35 of 40 subjects, whereas finger-cuff inflation scarcely disturbed the sleep. The performance of the ABPM 630 was excellent.

摘要

使用一种新型便携式动态血压监测设备(ABPM 630)获取的血压夜间行为,与通过手指容积示波装置(BP - 100)获取的进行了比较;前者使用由二氧化碳气体充气的传统臂式袖带以消除电动泵的噪音,并且基于袖带示波法以及柯氏音技术(麦克风法)。对于ABPM 630的麦克风法,与传统听诊法相比,收缩压的平均差值为 -0.28±6.15 mmHg(均值±标准差),舒张压为0.96±6.28 mmHg,并且ABPM 630测量的血压值与听诊法测量的血压值之间存在高度显著的相关性。在40例轻度至中度高血压患者中,使用ABPM 630和BP - 100同时监测24小时血压。ABPM 630测量的收缩压日间平均值与BP - 100测量的相似,而前者评估的夜间平均值显著高于后者评估的。ABPM 630的臂式袖带充气在40名受试者中的35名中引起了一定程度的睡眠干扰,而手指袖带充气几乎不干扰睡眠。ABPM 630的性能极佳。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验