Vet Rec. 2020 Oct 31;187(9):e70. doi: 10.1136/vr.105759. Epub 2020 May 15.
Cardiac arrhythmias are commonly auscultated during routine physical examinations in horses and determining the underlying electrical abnormality using an ECG is important. The most commonly used device is a three-lead base apex system (Televet), however few practitioners carry this for routine visits. With recognition of the utility of smartphone-based ECGs in humans, dogs and ruminants, the AliveCor single-lead bipolar smartphone-based ECG has gained popularity. The objective of this study was to determine if AliveCor and Televet ECG measurements were comparable in healthy horses using multiple observers.
ECGs were performed on 15 healthy horses simultaneously using the AliveCor and Televet.
There was very good to perfect interdevice and interobserver agreement for heart rate and RR interval measurement, and moderate-to-good interdevice and interobserver agreement for detection of non-pathological arrhythmias. Interdevice agreement for measurement of P-wave and QRS duration, QT, PR and T-peak to T-end interval was poor to fair. Interestingly, interobserver agreement for P-wave and QRS duration, QT, PR, and T-peak to T-end interval measurements was fair to good.
Overall, the AliveCor is comparable to the Televet for heart rate and RR measurement, and for the detection of non-pathogenic arrhythmias with acceptable agreement between observers.
在马的常规体检中,常可听诊到心律失常,使用心电图确定潜在的电异常非常重要。最常用的设备是三导联基底心尖系统(Televet),但很少有从业者在常规就诊时携带此设备。随着智能手机心电图在人类、狗和反刍动物中的应用得到认可,AliveCor 单导联双极智能手机心电图已广受欢迎。本研究的目的是确定在健康马中使用多个观察者时,AliveCor 和 Televet 心电图测量值是否具有可比性。
使用 AliveCor 和 Televet 同时对 15 匹健康马进行心电图检查。
心率和 RR 间期测量的设备间和观察者间一致性非常好到极好,非病理性心律失常的检测具有中度到良好的设备间和观察者间一致性。P 波和 QRS 时限、QT、PR 和 T 波峰至 T 波末间期的设备间一致性差到中等。有趣的是,P 波和 QRS 时限、QT、PR 和 T 波峰至 T 波末间期的观察者间一致性为中等至良好。
总体而言,AliveCor 与 Televet 相比,在心率和 RR 测量以及非致病性心律失常的检测方面具有可比性,观察者之间的一致性可接受。