Romanos Georgios E, Delgado-Ruiz Rafael Arcesio, Nicolas-Silvente Ana I
Periodontology, School of Dental Medicine, Stony Brook University, Stony Brook, NY 11794-8712, USA.
Prosthodontics and Digital Technology, School of Dental Medicine, Stony Brook University, Stony Brook, 11794-8712, NY, USA.
Materials (Basel). 2020 May 16;13(10):2306. doi: 10.3390/ma13102306.
The stability of the implant-abutment interface is crucial for the maintenance of the implant index integrity. Several factors are capable of inducing deformation in implant-abutment connection, such as the actual insertion of the implant into the bone. This study aimed to evaluate the deformations produced in the connection after the insertion of the implant. Ten implants with two different implant carriers (Type A: carrier attached to platform and Type B: carrier attached inside the index-connection) were placed in artificial Type II bone, and volumetric changes were evaluated for different connection features with a 3D digital microscope. ANOVA (analysis of variance), Wilcoxon, and Tukey HSD post-test were used for statistical comparisons. Type A implants presented deformation at the platform level (inner slot angles and slot width), but no volumetric changes were observed inside the connection. Type B implants presented deformation in three parameters inside the connection (outer channel length, coronal step width, and coronal step length). Within the limitations of this study, we can conclude that more deformation is expected at the internal connection when the implant carrier engages this area. The engagement area should be as far away as possible from the index connection.
种植体-基台界面的稳定性对于维持种植体指标完整性至关重要。多种因素能够导致种植体-基台连接发生变形,例如种植体实际植入骨内的操作。本研究旨在评估种植体植入后连接部位产生的变形情况。将10枚带有两种不同种植体载体的种植体(A型:载体附着于平台;B型:载体附着于指标连接内部)植入人工II型骨中,使用三维数字显微镜针对不同连接特征评估体积变化。采用方差分析(ANOVA)、威尔科克森检验以及图基HSD事后检验进行统计学比较。A型种植体在平台水平出现变形(内槽角度和槽宽度),但连接内部未观察到体积变化。B型种植体在连接内部的三个参数出现变形(外通道长度、冠部台阶宽度和冠部台阶长度)。在本研究的局限性范围内,我们可以得出结论,当种植体载体与该区域接触时,内部连接预计会出现更多变形。接触区域应尽可能远离指标连接。