The purpose of this in vitro study was to investigate the influence of eccentric loading on implant components by measuring screw loosening and observing these components under several load positions and magnitudes. The external and internal joint system implants with butt joint connection were subjected to cyclic loading tests according to the specifications of ISO 14801. Load position was set at 0, 4, or 8 mm, and load was set at 100 or 300 N. On the external joint system, the reverse torque values decreased with distal shift in the loading position and an increase in magnitude of load, and abrasion and deformation on the anti-rotation device were observed. On the internal joint system, no large decrease in reverse torque was observed even though increasing the load position and load, however, abrasion and deformation on the anti-rotation device as well as fracture at implant/abutment connection were observed.
本体外研究的目的是通过测量螺钉松动情况并在多个载荷位置和大小下观察植入部件,来研究偏心载荷对植入部件的影响。具有对接连接的外部和内部关节系统植入物按照ISO 14801的规范进行循环加载试验。载荷位置设定为0、4或8毫米,载荷设定为100或300牛。在外部关节系统上,反向扭矩值随着载荷位置向远端移动和载荷大小增加而降低,并且观察到防旋转装置上有磨损和变形。在内部关节系统上,即使增加载荷位置和载荷,反向扭矩也没有大幅下降,但是观察到防旋转装置上有磨损和变形以及植入物/基台连接处发生断裂。