Rubidium-82 generator yield and efficiency for PET perfusion imaging: Comparison of two clinical systems.

机构信息

Division of Cardiology, University of Ottawa Heart Institute, National Cardiac PET Centre, 40 Ruskin Street, Ottawa, ON, K1Y 4W7, Canada.

Division of Nuclear Medicine, The Ottawa Hospital, Ottawa, ON, Canada.

出版信息

J Nucl Cardiol. 2020 Oct;27(5):1728-1738. doi: 10.1007/s12350-020-02200-6. Epub 2020 May 20.

Abstract

INTRODUCTION

Strontium-82/Rubidium-82 (Sr/Rb) generators are used widely for positron emission tomography (PET) imaging of myocardial perfusion. In this study, the Rb isotope yield and production efficiency of two FDA-approved Sr/Rb generators were compared.

METHODS

N = 515 sequential daily quality assurance (QA) reports from 9 CardioGen-82 and 9 RUBY-FILL generators were reviewed over a period of 2 years. A series of test elutions was performed at different flow-rates on the RUBY-FILL system to determine an empirical correction-factor used to convert CardioGen-82 daily QA values of Rb activity (dose-calibrator 'maximum' of 50 mL elution at 50 mL·min) to RUBY-FILL equivalent values (integrated 'total' of 35 mL elution at 20 mL·min). The generator yield (Rb) and production efficiency (Rb yield/Sr parent activity) were measured and compared after this conversion to a common scale.

RESULTS

At the start of clinical use, the system reported Rb activity from daily QA was lower for CardioGen-82 vs RUBY-FILL (2.3 ± 0.2 vs 3.0 ± 0.2 GBq, P < 0.001) despite having similar Sr activity. Dose-calibrator 'maximum' (CardioGen-82) values were found to under-estimate the integrated 'total' (RUBY-FILL) activity by ~ 24% at 50 mL·min. When these data were used to convert the CardioGen-82 values to a common measurement scale (integrated total activity) the CardioGen-82 efficiency remained slightly lower than the RUBY-FILL system on average (88 ± 4% vs 95 ± 4%, P < 0.001). The efficiency of Rb production improved for both systems over the respective periods of clinical use.

CONCLUSIONS

Rb generator yield was significantly under-estimated using the CardioGen-82 vs RUBY-FILL daily QA procedure. When generator yield was expressed as the integrated total activity for both systems, the estimated Rb production efficiency of the CardioGen-82 system was ~ 7% lower than RUBY-FILL over the full period of clinical use.

摘要

简介

锶-82/铷-82(Sr/Rb)发生器广泛用于正电子发射断层扫描(PET)心肌灌注成像。在这项研究中,比较了两种获得美国食品药品监督管理局批准的 Sr/Rb 发生器的 Rb 同位素产率和生产效率。

方法

在两年的时间里,回顾了 9 个 CardioGen-82 和 9 个 RUBY-FILL 发生器的 515 份连续每日质量保证(QA)报告。在 RUBY-FILL 系统上以不同流速进行了一系列测试洗脱,以确定一个经验校正因子,用于将 CardioGen-82 每日 QA 的 Rb 活度值(洗脱体积为 50 mL 时剂量校准器“最大”值为 50 mL·min)转换为 RUBY-FILL 等效值(洗脱体积为 35 mL 时积分“总”值为 20 mL·min)。在转换为通用刻度后,测量并比较了发生器产率(Rb)和生产效率(Rb 产率/Sr 母核活度)。

结果

在临床应用开始时,尽管 Sr 活度相似,但 CardioGen-82 与 RUBY-FILL 相比,每日 QA 报告的 Rb 活度较低(2.3 ± 0.2 比 3.0 ± 0.2GBq,P < 0.001)。在 50 mL·min 时,剂量校准器“最大”(CardioGen-82)值被发现低估了积分“总”(RUBY-FILL)活性约 24%。当使用这些数据将 CardioGen-82 值转换为通用测量刻度(积分总活性)时,CardioGen-82 的效率仍略低于 RUBY-FILL 系统,平均为 88 ± 4%比 95 ± 4%,P < 0.001)。在各自的临床应用期间,两种系统的 Rb 生产效率均有所提高。

结论

使用 CardioGen-82 与 RUBY-FILL 每日 QA 程序,锶-82 发生器的产率被显著低估。当将发生器产率表示为两种系统的积分总活性时,CardioGen-82 系统的 Rb 生产效率在整个临床应用期间比 RUBY-FILL 系统低约 7%。