Strontium-82/Rubidium-82 (Sr/Rb) generators are used widely for positron emission tomography (PET) imaging of myocardial perfusion. In this study, the Rb isotope yield and production efficiency of two FDA-approved Sr/Rb generators were compared.
N = 515 sequential daily quality assurance (QA) reports from 9 CardioGen-82 and 9 RUBY-FILL generators were reviewed over a period of 2 years. A series of test elutions was performed at different flow-rates on the RUBY-FILL system to determine an empirical correction-factor used to convert CardioGen-82 daily QA values of Rb activity (dose-calibrator 'maximum' of 50 mL elution at 50 mL·min) to RUBY-FILL equivalent values (integrated 'total' of 35 mL elution at 20 mL·min). The generator yield (Rb) and production efficiency (Rb yield/Sr parent activity) were measured and compared after this conversion to a common scale.
At the start of clinical use, the system reported Rb activity from daily QA was lower for CardioGen-82 vs RUBY-FILL (2.3 ± 0.2 vs 3.0 ± 0.2 GBq, P < 0.001) despite having similar Sr activity. Dose-calibrator 'maximum' (CardioGen-82) values were found to under-estimate the integrated 'total' (RUBY-FILL) activity by ~ 24% at 50 mL·min. When these data were used to convert the CardioGen-82 values to a common measurement scale (integrated total activity) the CardioGen-82 efficiency remained slightly lower than the RUBY-FILL system on average (88 ± 4% vs 95 ± 4%, P < 0.001). The efficiency of Rb production improved for both systems over the respective periods of clinical use.
Rb generator yield was significantly under-estimated using the CardioGen-82 vs RUBY-FILL daily QA procedure. When generator yield was expressed as the integrated total activity for both systems, the estimated Rb production efficiency of the CardioGen-82 system was ~ 7% lower than RUBY-FILL over the full period of clinical use.
锶-82/铷-82(Sr/Rb)发生器广泛用于正电子发射断层扫描(PET)心肌灌注成像。在这项研究中,比较了两种获得美国食品药品监督管理局批准的 Sr/Rb 发生器的 Rb 同位素产率和生产效率。
在两年的时间里,回顾了 9 个 CardioGen-82 和 9 个 RUBY-FILL 发生器的 515 份连续每日质量保证(QA)报告。在 RUBY-FILL 系统上以不同流速进行了一系列测试洗脱,以确定一个经验校正因子,用于将 CardioGen-82 每日 QA 的 Rb 活度值(洗脱体积为 50 mL 时剂量校准器“最大”值为 50 mL·min)转换为 RUBY-FILL 等效值(洗脱体积为 35 mL 时积分“总”值为 20 mL·min)。在转换为通用刻度后,测量并比较了发生器产率(Rb)和生产效率(Rb 产率/Sr 母核活度)。
在临床应用开始时,尽管 Sr 活度相似,但 CardioGen-82 与 RUBY-FILL 相比,每日 QA 报告的 Rb 活度较低(2.3 ± 0.2 比 3.0 ± 0.2GBq,P < 0.001)。在 50 mL·min 时,剂量校准器“最大”(CardioGen-82)值被发现低估了积分“总”(RUBY-FILL)活性约 24%。当使用这些数据将 CardioGen-82 值转换为通用测量刻度(积分总活性)时,CardioGen-82 的效率仍略低于 RUBY-FILL 系统,平均为 88 ± 4%比 95 ± 4%,P < 0.001)。在各自的临床应用期间,两种系统的 Rb 生产效率均有所提高。
使用 CardioGen-82 与 RUBY-FILL 每日 QA 程序,锶-82 发生器的产率被显著低估。当将发生器产率表示为两种系统的积分总活性时,CardioGen-82 系统的 Rb 生产效率在整个临床应用期间比 RUBY-FILL 系统低约 7%。