BMC Ophthalmol. 2020 May 24;20(1):197. doi: 10.1186/s12886-020-01379-z.
Assessment of interobserver reproducibility and interocular symmetry using optical coherence tomography (OCT)-based measurements of the macular ganglion cell complex (GCC) in healthy children facilitates interpretation of OCT data. We assessed the interobserver reproducibility and interocular symmetry of GCC and evaluated candidate determinants.
This was a cross-sectional study performed in a primary and tertiary health-care setting. A total of 126 healthy participants aged 5 to 18 years were eligible. GCC scans were performed by 4 operators using the Topcon 3D OCT-2000 device. Intraclass correlation coefficients (ICCs) were used to estimate reproducibility and symmetry. Cut-off points for symmetry were defined as the 95th percentile of the absolute interocular difference for 6 GCC parameters. Percentile distributions of interocular difference were generated based on age and difference in absolute interocular spherical equivalent (SE).
The reproducibility ICC ranged from 0.96 to 0.98 for all 6 GCC parameters. Cut-off points for interocular symmetry of the superior and inferior quadrants and total macular retinal nerve fibre layer thickness (mRNFL) and macular ganglion cell layer-inner plexiform layer thickness were 3.5, 4.5, 3.0, 3.0, 2.5, and 2.5 μm respectively. A positive association was observed between the absolute interocular difference of SE and superior and total mRNFL symmetry values (p = 0.047 and p = 0.040, respectively).
OCT measurements of GCC in healthy children show excellent reproducibility. Interocular differences in SE should be assessed when mRNFL differences exceed the 95% cut-off. These findings can contribute to establish reference values for interocular symmetry in paediatric GCC parameters.
使用基于光学相干断层扫描(OCT)的黄斑神经节细胞复合体(GCC)测量值评估健康儿童的观察者间可重复性和眼间对称性,有助于解释 OCT 数据。我们评估了 GCC 的观察者间可重复性和眼间对称性,并评估了候选决定因素。
这是一项在初级和三级保健环境中进行的横断面研究。共有 126 名年龄在 5 至 18 岁的健康参与者符合条件。GCC 扫描由 4 名操作人员使用 Topcon 3D OCT-2000 设备进行。使用组内相关系数(ICC)来估计可重复性和对称性。对称性的截断值定义为 6 个 GCC 参数的眼间绝对差异的第 95 个百分位数。根据年龄和眼间绝对等效球镜(SE)差异生成眼间差异的百分位分布。
所有 6 个 GCC 参数的可重复性 ICC 范围为 0.96 至 0.98。上象限和下象限以及总黄斑视网膜神经纤维层厚度(mRNFL)和黄斑神经节细胞层-内丛状层厚度的眼间对称性的截断值分别为 3.5、4.5、3.0、3.0、2.5 和 2.5μm。SE 的眼间绝对差异与上象限和总 mRNFL 对称性值呈正相关(p=0.047 和 p=0.040)。
健康儿童的 GCC OCT 测量值显示出极好的可重复性。当 mRNFL 差异超过 95%的截断值时,应评估 SE 的眼间差异。这些发现有助于建立儿童 GCC 参数眼间对称性的参考值。