Piłat Paweł, Borzęcki Adam, Jazienicki Mieczysław, Gerkowicz Agnieszka, Szubstarski Franciszek, Krasowska Dorota
Chair and Department of Dermatology, Venereology and Pediatric Dermatology, Medical University of Lublin, Lublin, Poland.
Non-Public Health Care Center Med-Laser, Lublin, Poland.
Postepy Dermatol Alergol. 2020 Apr;37(2):207-213. doi: 10.5114/ada.2018.79939. Epub 2019 Jan 21.
There are many papers in the available literature on the pre-operative evaluation of malignant melanoma (MM) using a 20 MHz probe. In many Western European countries, this examination is standard in the diagnosis of such lesions in terms of the resection margin. However, few papers describe and compare the results of melanoma thickness measurements using 20 and 50 MHz frequency probes.
The aim of the study was to evaluate the thickness of melanoma using 20 and 50 MHz probes. In addition, the obtained values were evaluated in comparison to the thickness values obtained in the histopathological examination.
Melanoma malignum was confirmed histopathologically in 19 out of 72 patients. Then only those 19 patients were enrolled in the study.
A correlation was demonstrated between the thickness of the melanoma measured with the 20 and 50 MHz probes, expressed in mm, and the thickness obtained in the histopathological examination expressed in millimeters (Breslow's scale). It was 0.759 ( < 0.001) and 0.734 ( < 0.001), respectively. The average thickness of melanomas obtained in ultrasound examination was identical for both probes and was 0.74 mm.
It seems that both types of probes can be used to assess melanoma thickness. The demonstrated correlation of high-frequency ultrasonography (HFUS) thickness with Breslow's scale indicates that the HFUS evaluation can be used in determination of melanoma resection margins as a repeatable, painless, non-invasive test.
现有文献中有许多关于使用20兆赫探头对恶性黑色素瘤(MM)进行术前评估的论文。在许多西欧国家,就切缘而言,这项检查是此类病变诊断的标准方法。然而,很少有论文描述和比较使用20兆赫和50兆赫频率探头测量黑色素瘤厚度的结果。
本研究的目的是使用20兆赫和50兆赫探头评估黑色素瘤的厚度。此外,将获得的值与组织病理学检查中获得的厚度值进行比较评估。
72例患者中有19例经组织病理学确诊为恶性黑色素瘤。然后仅将这19例患者纳入研究。
以毫米为单位表示的用20兆赫和50兆赫探头测量的黑色素瘤厚度与以毫米( Breslow量表)表示的组织病理学检查中获得的厚度之间存在相关性。分别为0.759(<0.001)和0.734(<0.001)。两种探头在超声检查中获得的黑色素瘤平均厚度相同,均为0.74毫米。
似乎两种类型的探头都可用于评估黑色素瘤厚度。高频超声(HFUS)厚度与Breslow量表之间已证实的相关性表明,HFUS评估可作为一种可重复、无痛、非侵入性的检查用于确定黑色素瘤的切缘。