Romero-Franco Natalia, Jiménez-Reyes Pedro, González-Hernández Jorge Miguel, Fernández-Domínguez Juan Carlos
Nursing and Physiotherapy Department, University of the Balearic Islands, E-07122, Palma de Mallorca, Spain.
Centre for Sport Studies, Rey Juan Carlos University, E-28943, Madrid, Spain.
Phys Ther Sport. 2020 Jul;44:136-142. doi: 10.1016/j.ptsp.2020.05.003. Epub 2020 May 16.
To analyze the concurrent validity and reliability of an iPhone application for assessing range of motion (ROM) and joint position sense (JPS) in ankle and knee joints.
Cross-sectional study.
Sport laboratory.
Twenty healthy and physically active volunteers.
All participants performed a ROM and a JPS test in ankle and knee joints, which were simultaneously evaluated with photo-analysis and the iPhone application.
A total of 80 angles were obtained with the iPhone app and the photo-analysis and compared for concurrent validity. Reliability was evaluated through re-scoring of images with the iPhone app by two different testers.
Very high correlation was observed between both methods for ankle and knee ROM and knee JPS (r > 0.90), and high correlation for ankle JPS (r = 0.71-0.90), while Bland-Altman plots showed absolute agreement for all the variables. Inter- and intra-tester reliability was perfect for all the variables (ICC > 0.81), except for the inter-tester reliability of ankle JPS, which was substantial (ICC = 0.61-0.81).
This new iPhone application is valid and reliable for measuring ankle and knee ROM and JPS, although special attention is needed during ankle evaluation to avoid errors.
分析一款用于评估踝关节和膝关节活动范围(ROM)及关节位置觉(JPS)的iPhone应用程序的同时效度和信度。
横断面研究。
运动实验室。
20名健康且身体活跃的志愿者。
所有参与者在踝关节和膝关节进行ROM和JPS测试,同时使用照片分析和iPhone应用程序进行评估。
通过iPhone应用程序和照片分析共获得80个角度,并对其同时效度进行比较。通过两名不同测试者使用iPhone应用程序对图像重新评分来评估信度。
两种方法在踝关节和膝关节ROM以及膝关节JPS方面观察到非常高的相关性(r>0.90),在踝关节JPS方面观察到高相关性(r = 0.71 - 0.90),而Bland-Altman图显示所有变量的绝对一致性。除踝关节JPS的测试者间信度为中等(ICC = 0.61 - 0.81)外,所有变量的测试者间和测试者内信度均极佳(ICC>0.81)。
这款新的iPhone应用程序在测量踝关节和膝关节ROM及JPS方面是有效且可靠的,不过在踝关节评估过程中需要特别注意以避免误差。