健康志愿者脑血管自动调节的组织氧合指数:两种近红外光谱设备的比较

Tissue oxygenation indices of cerebrovascular autoregulation in healthy volunteers: a comparison of two NIRS devices.

作者信息

Peng Alexander, Chua Matthew J, Chan Bernard, Jaeger Matthias, Aneman Anders, Chuan Alwin

机构信息

Department of Intensive Care, Blacktown Hospital , Sydney, Australia.

Department of Intensive Care, Nepean Hospital , Sydney, Australia.

出版信息

Neurol Res. 2020 Oct;42(10):897-903. doi: 10.1080/01616412.2020.1790869. Epub 2020 Jul 9.

Abstract

BACKGROUND

Correlation coefficients between blood pressure and cerebral oxygen saturation measured using near-infrared spectrometry may be used to derive the tissue oximetry index of cerebral autoregulation. Cerebral oxygen saturations demonstrate poor agreement between near-infrared spectrometers however it is unclear if measurements of autoregulation are similarly specific to the equipment used.

METHODS

Cerebral oxygen saturation was monitored bilaterally in 74 healthy volunteers using both the FORE-SIGHT and EQUANOX monitors in random order. The tissue oximetry index was calculated during changes in blood pressure induced by isometric handgrip manoeuvres and the mean bias and limits of agreement were calculated.

RESULTS

Tissue oximetry index measured by FORE-SIGHT was higher than EQUANOX (0.21 ± 0.16 versus 0.15 ± 0.17, P < 0.001) and limits of agreement were -0.24 to 0.36. Baseline cerebral oxygen saturation by FORE-SIGHT was lower than EQUANOX by 1.48% (CI95% 0.63-2.33) and limits of agreement ranged from -11.8% to 8.8%.

CONCLUSIONS

The substantial bias and wide limits of agreement for the tissue oximetry index as a measure of cerebral autoregulation indicate that values must be treated as specific to models of near-infrared spectrometers.

摘要

背景

使用近红外光谱法测量的血压与脑氧饱和度之间的相关系数可用于推导脑自动调节的组织氧饱和度指数。然而,近红外光谱仪之间的脑氧饱和度显示出较差的一致性,目前尚不清楚自动调节的测量对于所用设备是否同样具有特异性。

方法

74名健康志愿者使用FORE-SIGHT和EQUANOX监测仪以随机顺序双侧监测脑氧饱和度。在等长握力动作引起血压变化期间计算组织氧饱和度指数,并计算平均偏差和一致性界限。

结果

FORE-SIGHT测量的组织氧饱和度指数高于EQUANOX(0.21±0.16对0.15±0.17,P<0.001),一致性界限为-0.24至0.36。FORE-SIGHT测量的基线脑氧饱和度比EQUANOX低1.48%(CI95%0.63-2.33),一致性界限范围为-11.8%至8.8%。

结论

作为脑自动调节测量指标的组织氧饱和度指数存在显著偏差和较宽的一致性界限,这表明其值必须被视为特定于近红外光谱仪型号。