From the Department of Nephrology, Internal Medicine, Hallym University Sacred Heart Hospital, Gyeonggi-do, Republic of Korea.
Simul Healthc. 2021 Feb 1;16(1):73-77. doi: 10.1097/SIH.0000000000000469.
This study was conducted to analyze the effectiveness of a custom-made Doppler ultrasound (DUS) flow simulator, vascular phantom, and Doppler test fluid in the training of dialysis staff in flow volume (FV) measurements for arteriovenous (AV) access in hemodialysis (HD) patients.
A DUS flow simulator was constructed using a continuous renal replacement therapy machine. Vascular phantoms were constructed using a rubber enema tube and keyboard cleaning gel. The Doppler test fluid consisted of freeze-dried instant coffee granules and 0.9% saline. This easy and affordable simulator was used to train 12 dialysis staff members, with no previous experience in DUS FV measurement. After a 3-day lecture course, the staff members performed DUS analyses on the AV access of HD patients. Thereafter, they underwent a 3-day training course using the simulator and then performed the DUS analyses of the AV access again. Each staff member assessed the FV 3 times, and the mean values of the measurements pretraining and posttraining were analyzed using paired t tests.
The difference in the mean value of FV measurements and the reference value decreased from 131.6 mL/min to 62.5 mL/min (95% confidence interval = 30.0-108.0, P = 0.002), and the standard deviation of the FV measurements decreased from 96.9 mL/min to 47.0 mL/min (95% confidence interval = 7.9-91.8, P = 0.023) after DUS training with the simulator.
The accuracy and reproducibility of FV measurements markedly improved after training with the simulator; it may be helpful for medical practitioners involved in AV access for HD treatment.
本研究旨在分析定制的多普勒超声(DUS)流量模拟器、血管模型和多普勒测试液在培训血液透析(HD)患者动静脉(AV)通路流量(FV)测量的透析工作人员方面的效果。
使用连续肾脏替代治疗机构建 DUS 流量模拟器。血管模型使用橡胶灌肠管和键盘清洁凝胶构建。多普勒测试液由冻干速溶咖啡颗粒和 0.9%生理盐水组成。使用这种简单且经济实惠的模拟器对 12 名没有 DUS FV 测量经验的透析工作人员进行培训。在为期 3 天的讲座课程之后,工作人员对 HD 患者的 AV 通路进行了 DUS 分析。此后,他们使用模拟器进行了为期 3 天的培训课程,然后再次对 AV 通路进行了 DUS 分析。每位工作人员评估了 FV 3 次,使用配对 t 检验分析了培训前后测量的平均值。
FV 测量的平均值与参考值的差异从 131.6 mL/min 减少到 62.5 mL/min(95%置信区间=30.0-108.0,P=0.002),FV 测量的标准差从 96.9 mL/min 减少到 47.0 mL/min(95%置信区间=7.9-91.8,P=0.023)。
使用模拟器进行 DUS 培训后,FV 测量的准确性和可重复性显著提高,可能有助于从事 HD 治疗的 AV 通路的医疗从业者。