Suppr超能文献

呼气气流阻力对呼吸吸气功的影响。

Effects of expiratory flow resistance on inspiratory work of breathing.

作者信息

Banner M J, Downs J B, Kirby R R, Smith R A, Boysen P G, Lampotang S

机构信息

Department of Anesthesiology, College of Medicine, University of Florida, Gainesville 32610-0254.

出版信息

Chest. 1988 Apr;93(4):795-9. doi: 10.1378/chest.93.4.795.

Abstract

To minimize work of breathing, airway pressure should not fluctuate during spontaneous breathing with continuous positive airway pressure (CPAP). However, flow resistance in the inspiratory limb of the breathing circuit and an inadequate continuous gas flow rate result in airway pressure fluctuation and increased work of breathing. Flow resistance of the expiratory pressure/exhalation valve also directly affects the level of airway pressure during spontaneous inhalation with CPAP (the greater the resistance of the valve, the greater the decrease in airway pressure and work of breathing). We compared this effect with three types of expiratory pressure valves: a threshold resistor with low resistance to flow, an inflatable balloon (mushroom) valve with moderate resistance to flow, and a variable-orifice flow resistor with a high resistance to flow. Work increased up to threefold with the balloon valve and more than tenfold with the flow resistor compared with the threshold resistor. To apply CPAP, expiratory pressure valves with low resistance to flow should be used to minimize fluctuations in airway pressure and, thus, in the work of spontaneous breathing.

摘要

为使呼吸功最小化,在持续气道正压通气(CPAP)下自主呼吸时气道压力不应波动。然而,呼吸回路吸气支的气流阻力以及持续气流速率不足会导致气道压力波动并增加呼吸功。呼气压力/呼气阀的气流阻力也直接影响CPAP自主吸气时的气道压力水平(阀门阻力越大,气道压力下降和呼吸功增加越明显)。我们比较了三种呼气压力阀的这种影响:一种对气流阻力低的阈值电阻器、一种对气流阻力适中的可充气气球(蘑菇)阀和一种对气流阻力高的可变孔口气流电阻器。与阈值电阻器相比,使用气球阀时呼吸功增加了三倍,使用气流电阻器时增加了十倍以上。为应用CPAP,应使用对气流阻力低的呼气压力阀,以尽量减少气道压力波动,从而减少自主呼吸功。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验