Work of breathing, inspiratory flow response, and expiratory resistance during continuous positive airway pressure with the ventilators EVITA-2, EVITA-4 and SV 300.

作者信息

Calzia E, Lindner K H, Stahl W, Martin A, Radermacher P, Georgieff M

机构信息

Department of Anesthesiology, University Hospital Ulm, Germany.

出版信息

Intensive Care Med. 1998 Sep;24(9):931-8. doi: 10.1007/s001340050692.

Abstract

OBJECTIVE

To analyze work of breathing (WOB) imposed by the respirators EVITA-2, EVITA-4 (Drägerwerk, Lübeck, Germany) and SV 300 (Siemens-Elema, Sweden) as well as inspiratory flow response and expiratory flow resistance during continuous positive airway pressure (CPAP).

DESIGN

Five study conditions on a lung model (CPAP at 0, 5, and 10 mbar, CPAP 5 mbar plus pressure support 2 mbar with both EVITA models, and CPAP 5 mbar with decreasing levels of flow and pressure trigger sensitivity with the SV 300) and three randomized study conditions in nine patients recovering from open heart surgery (condition A: EVITA-2, CPAP 5 mbar; condition B: SV 300, CPAP 5 mbar, flow trigger; condition C: SV 300, pressure trigger-4 mbar).

SETTING

University hospital intensive care unit and laboratory of pulmonary physiology.

MEASUREMENTS AND RESULTS

At each study condition we measured WOB, pressure-time product (PTP), WOB and PTP imposed (WOBimposed and PTPimposed), tidal volume, minute ventilation, respiratory rate, inspiratory trigger time, trigger pressure, trigger PTP, duration of inspiration, mean and peak inspiratory flow, and the delay from the onset of inspiration to peak inspiratory flow. Since the SV 300 automatically generates an additional pressure support of 2 cm H2O PTP, WOB, WOBimposed, and PTPimposed were higher with the EVITA-2 and EVITA-4 regardless of the trigger sensitivity set on the SV 300. The difference was neutralized with both types of EVITA ventilator by adding 2 mbar of pressure support during CPAP in order to achieve comparable conditions. Inspiratory flow response was faster with both EVITA models, expiratory flow resistance was higher with the SV 300. Decrements of trigger sensitivity with the SV 300 accelerated the flow response.

CONCLUSIONS

Under similar conditions, no difference in WOBimposed was observed, although inspiratory flow response and expiratory flow resistance differed substantially between the three ventilators tested. Trigger sensitivity plays a minor role in determining PTP and WOB but has major influence on flow.

摘要

目的

分析德国吕贝克德尔格公司生产的EVITA - 2、EVITA - 4呼吸机以及瑞典西门子 - 伊莱玛公司生产的SV 300呼吸机在持续气道正压通气(CPAP)期间施加的呼吸功(WOB)以及吸气流量响应和呼气流量阻力。

设计

在肺模型上设置五种研究条件(0、5和10 mbar的CPAP,两种EVITA型号在5 mbar CPAP基础上加2 mbar压力支持,以及SV 300在5 mbar CPAP下设置不同水平的流量和压力触发灵敏度),并在9例心脏直视手术后恢复的患者中设置三种随机研究条件(A组:EVITA - 2,5 mbar CPAP;B组:SV 300,5 mbar CPAP,流量触发;C组:SV 300,压力触发 - 4 mbar)。

设置

大学医院重症监护病房和肺生理学实验室。

测量与结果

在每个研究条件下,我们测量了WOB、压力 - 时间乘积(PTP)、施加的WOB和PTP(WOBimposed和PTPimposed)、潮气量、分钟通气量、呼吸频率、吸气触发时间、触发压力、触发PTP、吸气持续时间、平均和峰值吸气流量以及从吸气开始到峰值吸气流量的延迟。由于SV 300会自动产生2 cm H2O PTP的额外压力支持,无论在SV 300上设置何种触发灵敏度,EVITA - 2和EVITA - 4的WOB、WOBimposed和PTPimposed都更高。在CPAP期间通过添加2 mbar压力支持使两种EVITA呼吸机达到可比条件,从而消除了差异。两种EVITA型号的吸气流量响应更快,SV 300的呼气流量阻力更高。降低SV 300的触发灵敏度可加快流量响应。

结论

在相似条件下,尽管测试的三种呼吸机之间吸气流量响应和呼气流量阻力存在显著差异,但未观察到施加的WOB有差异。触发灵敏度在确定PTP和WOB方面作用较小,但对流量有重大影响。