Qi Peng, Li Xue-Xun, Tian Ying, Shi Liang, Wang Yan-Jiang, Liu Xing-Peng
Heart Center, Beijing Chao-Yang Hospital, Capital Medical University, Beijing, China.
Department of Cardiology, Hebei General Hospital, Shijiazhuang, Hebei, China.
J Int Med Res. 2020 Aug;48(8):300060520947880. doi: 10.1177/0300060520947880.
A 70-year-old woman with symptomatic bradycardia caused by persistent atrial fibrillation and atrioventricular block was referred to our institution for pacemaker implantation. After we failed to obtain adequate His bundle capture thresholds (>2.5 V at 1.0 ms) at three pacing sites, left bundle branch pacing was attempted as an alternative technique. The tip of the 3830 lead was screwed towards the left side of the interventricular septum. Contrast medium was injected through the C315 sheath, which was placed close to the right side of the interventricular septum to determine the exact depth of the 3830 lead inside the septum. Unexpectedly, the vessels in the interventricular septum were revealed by the contrast, which showed that the lead had penetrated one of the septal vessels. To the best of our knowledge, this is the first reported case of a patient in whom injection of a contrast agent through a delivery sheath showed damage to the interventricular septal vessels. Findings from this case suggest that injection of contrast medium through a C315 sheath that is placed close to the interventricular septum is a potential method for excluding damage to interventricular septal vessels.
一名70岁女性因持续性心房颤动和房室传导阻滞导致症状性心动过缓,被转诊至我院行起搏器植入术。在三个起搏部位均未能获得足够的希氏束夺获阈值(1.0毫秒时>2.5伏)后,尝试采用左束支起搏作为替代技术。将3830导线的尖端拧向室间隔左侧。通过置于室间隔右侧附近的C315鞘管注入造影剂,以确定3830导线在间隔内的确切深度。出乎意料的是,造影显示室间隔内的血管,表明导线已穿透其中一根间隔血管。据我们所知,这是首例通过输送鞘管注射造影剂显示室间隔血管受损的病例报告。该病例的发现表明,通过置于室间隔附近的C315鞘管注射造影剂是排除室间隔血管损伤的一种潜在方法。