J Refract Surg. 2020 Sep 1;36(9):625-632. doi: 10.3928/1081597X-20200710-01.
To test the performance of a new monofocal intraocular lens, intended to extend depth of focus (Tecnis Eyhance, ICB00; Johnson & Johnson Vision, Inc) (ICB-IOL), in comparison to a standard monofocal IOL (Tecnis 1-piece, ZCB00; Johnson & Johnson Vision, Inc) (ZCB-IOL) of the same platform and material.
Assessment of the optical performance of the two IOLs was made in vitro using an optical test bench with a model eye. The spherical aberration, modulation transfer function (MTF), and area under the MTF (MTFa) were obtained for pupil sizes ranging from 2 to 5 mm. Through-focus MTFa curves between -3.00 and +1.00 diopters (D) were obtained with three pupil sizes (2, 3, and 4.5 mm). Halo formation was also assessed for both lenses.
The ICB-IOL had slightly worse optical quality at its best focus (ie, lower MTF scores at distance vision) and more negative spherical aberration than the ZCB-IOL for pupils ranging from 2 to 3 mm. The maximum of the through-focus MTFa curve of the ICB-IOL with a 2-mm pupil shifted to a myopic defocus of -0.50 D. For larger pupils (≥ 3.5 mm), there were no differences in spherical aberration, MTF scores, and halo energy between the two lenses.
The new ICB-IOL is a modified monofocal lens with 0.50 D of additional power in its central 2-mm zone and more negative spherical aberration values, which induce a myopic shift of the maximum of optical quality and could improve intermediate vision. For pupils larger than 3.5 mm, there were no differences between IOLs. The new ICB-IOL design would produce photic phenomena comparable to a standard IOL. [J Refract Surg. 2020;36(9):625-632.].
测试一种新型单焦点人工晶状体(旨在扩展景深的 Tecnis Eyhance,ICB00;Johnson & Johnson Vision, Inc)(ICB-IOL)的性能,与相同平台和材料的标准单焦点 IOL(Tecnis 1-piece,ZCB00;Johnson & Johnson Vision, Inc)(ZCB-IOL)进行比较。
使用带有模型眼的光学测试台对两种 IOL 的光学性能进行了评估。获得了瞳孔尺寸为 2 至 5 毫米的情况下的球差、调制传递函数(MTF)和 MTF 下面积(MTFa)。获得了三个瞳孔尺寸(2、3 和 4.5 毫米)下-3.00 至+1.00 屈光度(D)之间的焦点转移 MTFa 曲线。还评估了两种镜片的晕影形成。
在最佳焦点(即距离视力时较低的 MTF 分数)下,ICB-IOL 的光学质量略差,并且对于 2 至 3 毫米的瞳孔,ICB-IOL 的球差更负。具有 2 毫米瞳孔的 ICB-IOL 的焦点转移 MTFa 曲线的最大值向近视离焦-0.50 D 移动。对于较大的瞳孔(≥3.5 毫米),两种镜片之间的球差、MTF 分数和晕影能量没有差异。
新型 ICB-IOL 是一种改良的单焦点镜片,其中心 2 毫米区域的附加功率为 0.50 D,球差值更负,这会导致最大光学质量的近视移位,并可改善中间视力。对于大于 3.5 毫米的瞳孔,两种人工晶状体之间没有差异。新的 ICB-IOL 设计会产生与标准 IOL 相当的光学现象。[J Refract Surg. 2020;36(9):625-632.]。