Evaluation of a shorter algorithm in an automated analysis of sublingual microcirculation.

机构信息

Department of Cardiology, Pulmonary Diseases, and Vascular Medicine, Medical Faculty, University Hospital Düsseldorf, Düsseldorf, Germany.

Department of Anesthesiology and Critical Care, Medical Center - University of Freiburg, Faculty of Medicine, University of Freiburg, Freiburg, Germany.

出版信息

Clin Hemorheol Microcirc. 2020;76(2):287-297. doi: 10.3233/CH-209201.

Abstract

OBJECTIVE

Diagnostic and risk stratification in intensive and emergency medicine must be fast, accurate, and reliable. The assessment of sublingual microcirculation is a promising tool for this purpose. However, its value is limited because the measurement is time-consuming in unstable patients. This proof-of-concept validation study examines the non-inferiority of a reduced frame rate in image acquisition regarding quality, measurement results, and time.

METHODS

This prospective observational study included healthy volunteers. Sublingual measurement of microcirculation was performed using a sidestream dark field camera (SDF, MicroVision Medical®). Video-quality was evaluated with a modified MIQS (microcirculation image quality score). AVA 4.3C software calculated microcirculatory parameters.

RESULTS

Thirty-one volunteers were included. There was no impact of the frame rate on the time needed by the software algorithm to measure one video (4.5 ± 0.5 minutes) for AVA 4.3C. 86 frames per video provided non inferior video quality (MIQS 1.8 ± 0.7 for 86 frames versus MIQS 2.2 ± 0.6 for 215 frames, p < 0.05), equal results for all microcirculatory parameters, but did not result in an advantage in terms of speed. No complications occurred.

CONCLUSION

Video captures with 86 frames offer equal video quality and results for consensus parameters compared to 215 frames. However, there was no advantage regarding the time needed for the overall measurement procedure.

摘要

目的

在重症和急救医学中,诊断和风险分层必须快速、准确、可靠。评估舌下微循环是一种很有前途的工具。然而,由于在不稳定的患者中测量时间较长,其价值有限。本概念验证研究旨在检验图像采集帧率降低对质量、测量结果和时间的非劣效性。

方法

本前瞻性观察研究纳入了健康志愿者。使用侧流暗场相机(SDF,MicroVision Medical®)进行舌下微循环测量。使用改良的 MIQS(微循环图像质量评分)评估视频质量。AVA 4.3C 软件计算微循环参数。

结果

31 名志愿者纳入研究。帧率对 AVA 4.3C 测量一个视频所需的软件算法时间(4.5±0.5 分钟)没有影响。每视频 86 帧提供了非劣视频质量(86 帧的 MIQS 为 1.8±0.7,215 帧的 MIQS 为 2.2±0.6,p<0.05),所有微循环参数的结果相同,但在速度方面没有优势。未发生并发症。

结论

与 215 帧相比,86 帧的视频采集在共识参数方面提供了相等的视频质量和结果。然而,在整个测量过程所需的时间方面没有优势。