Suppr超能文献

对经需氧处理的血清进行pH调节后,血清中离子钙的离子选择性测量新技术。

New techniques for ion-selective measurements of ionized calcium in serum after pH adjustment of aerobically handled sera.

作者信息

Schwartz H D

出版信息

Clin Chem. 1976 Apr;22(4):461-7.

PMID:3297
Abstract

I report further experience in measuring ionized calcium (Ca2+) with the AMT Electron System and its serum standards and solid-state, dip, calcium-selective electrodes. With this system, serum pH can be adjusted with CO2 gas and Ca2+ and pH simultaneously measured; when 5.2% CO2 (40 mm pco2) is used for sample equilibration, the standard bicarbonate concentration is also provided. I measured serum Ca2+ as a function of pH between pH 7.0 and 9.0 and found the relationship to be reproducible, with no evidence of irreversible complexing of Ca2+. When the pH of aerobically exposed, mailed sera was restored to the original values, their values for Ca2+ were the same as for the fresh sera. Measurement of Ca2+ in routinely (aerobically) handled sera after pH restoration with CO2 gas was therefore validated, both samples from within an institution and mailed specimens. Standardization to pH 7.40 is recommended for routine measurements, is generally more accurate than use of heparin or quasianaerobic techniques, and is a practical approach. In patients with possible uncompensated acid-base disturbance (which may be indicated by an abnormal standard bicarbonate concentration if not suspected clinically), patient pH should be measured independently as part of the usual strict, anaerobic blood-gas-analysis procedures. Abnormal patient pH must be considered in the interpretation of Ca2+ results determined at pH 7.40 which are borderline or slightly abnormal; most accurately, Ca2+ may be measured in the separated sera at the previously determined patient pH value. Studies of aqueous solutions with the currently used Ca2+ electrodes showed a selectivity coefficient (the constant which relates the activity of an interfering ion to the activity of calcium that would contribute the same emf) KNa=0.0031 +/- 0.0003 (SE) and KMg=0.046 +/- 0.004 (SE). At physiological concentrations of Ca2+, physiologically encountered variation in Na+ is of no significance in resulting Ca2+, but extreme variation in Mg2+ may cause an error of approximately 1%.

摘要

我报告了使用AMT电子系统及其血清标准品和固态浸入式钙选择性电极测量离子钙(Ca2+)的更多经验。使用该系统时,可通过二氧化碳气体调节血清pH值,并同时测量Ca2+和pH值;当使用5.2%的二氧化碳(40毫米汞柱pco2)进行样品平衡时,还可提供标准碳酸氢盐浓度。我测量了pH值在7.0至9.0之间时血清Ca2+随pH值的变化情况,发现这种关系具有可重复性,没有Ca2+发生不可逆络合的证据。当有氧暴露并邮寄的血清pH值恢复到原始值时,其Ca2+值与新鲜血清相同。因此,在用二氧化碳气体恢复pH值后,对常规(有氧)处理的血清中Ca2+的测量得到了验证,无论是来自同一机构的样本还是邮寄的标本。建议常规测量时将pH值标准化为7.40,这通常比使用肝素或准厌氧技术更准确,且是一种实用的方法。对于可能存在未代偿酸碱紊乱的患者(如果临床未怀疑,可能表现为标准碳酸氢盐浓度异常),应按照常规严格的厌氧血气分析程序独立测量患者的pH值。在解释pH值为7.40时测定的Ca2+结果处于临界值或略异常时,必须考虑患者异常的pH值;最准确的做法是在先前测定的患者pH值下测量分离血清中的Ca2+。使用当前的Ca2+电极对水溶液进行的研究表明,选择性系数(将干扰离子的活度与产生相同电动势的钙的活度相关联的常数)KNa = 0.0031 ± 0.0003(标准误差),KMg = 0.046 ± 0.004(标准误差)。在Ca2+的生理浓度下,生理上遇到的Na+变化对所得Ca2+无显著影响,但Mg2+的极端变化可能导致约1%的误差。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验