测试环境和加载速率对椎间盘压缩力学的影响:三个不同实验室的重复性评估

Influence of testing environment and loading rate on intervertebral disc compressive mechanics: An assessment of repeatability at three different laboratories.

作者信息

Newell Nicolas, Rivera Tapia David, Rahman Tamanna, Lim Shiyin, O'Connell Grace D, Holsgrove Timothy P

机构信息

Department of Mechanical Engineering Imperial College London London UK.

Department of Engineering University of Exeter Exeter UK.

出版信息

JOR Spine. 2020 Jul 15;3(3):e21110. doi: 10.1002/jsp2.1110. eCollection 2020 Sep.

Abstract

In vitro mechanical testing of intervertebral discs is crucial for basic science and pre-clinical testing. Generally, these tests aim to replicate in vivo conditions, but simplifications are necessary in specimen preparation and mechanical testing due to complexities in both structure and the loading conditions required to replicate in vivo conditions. There has been a growing interest in developing a consensus of testing protocols within the spine community to improve comparison of results between studies. The objective of this study was to perform axial compression experiments on bovine bone-disc-bone specimens at three institutions. No differences were observed between testing environment being air, with PBS soaked gauze, or a PBS bath ( > .206). A 100-fold increase in loading rate resulted in a small (2%) but significant increase in compressive mechanics ( < .017). A 7% difference in compressive stiffness between Labs B and C was eliminated when values were adjusted for test system compliance. Specimens tested at Lab A, however, were found to be stiffer than specimens from Lab B and C. Even after normalizing for disc geometry and adjusting for system compliance, an ∼35% difference was observed between UK based labs (B and C) and the USA based lab (A). Large differences in specimen stiffness may be due to genetic differences between breeds or in agricultural feed and use of growth hormones; highlighting significant challenges in comparing mechanics data across studies. This research provides a standardized test protocol for the comparison of spinal specimens and provides steps towards understanding how location and test set-up may affect biomechanical results.

摘要

椎间盘的体外力学测试对于基础科学和临床前测试至关重要。一般来说,这些测试旨在复制体内条件,但由于结构的复杂性以及复制体内条件所需的加载条件的复杂性,在样本制备和力学测试中进行简化是必要的。脊柱领域越来越关注制定测试方案的共识,以改善不同研究结果之间的比较。本研究的目的是在三个机构对牛骨 - 椎间盘 - 骨标本进行轴向压缩实验。在空气环境、用 PBS 浸泡纱布或 PBS 浴的测试环境之间未观察到差异(>.206)。加载速率提高 100 倍导致压缩力学有小幅(2%)但显著的增加(<.017)。当对测试系统的顺应性进行值调整时,实验室 B 和 C 之间压缩刚度 7% 的差异被消除。然而,发现实验室 A 测试的标本比实验室 B 和 C 的标本更硬。即使对椎间盘几何形状进行归一化并对系统顺应性进行调整后,英国的实验室(B 和 C)与美国的实验室(A)之间仍观察到约 35% 的差异。标本刚度的巨大差异可能是由于品种之间的遗传差异、农业饲料或生长激素的使用;这突出了跨研究比较力学数据的重大挑战。本研究为脊柱标本的比较提供了标准化测试方案,并为理解位置和测试设置如何影响生物力学结果提供了步骤。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/81e2/7524251/4f81b6f5f517/JSP2-3-e21110-g001.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验