Viruses. 2020 Oct 20;12(10):1184. doi: 10.3390/v12101184.
RT-PCR on nasopharyngeal (NPS)/oropharyngeal swabs is the gold standard for diagnosis of SARS-CoV-2 infection and viral load monitoring. Oral fluid (OF) is an alternate clinical sample, easy and safer to collect and could be useful for COVID-19 diagnosis, monitoring viral load and shedding.
Optimal assay conditions and analytical sensitivity were established for the commercial Simplexa™ COVID-19 Direct assay adapted to OF matrix. The assay was used to test 337 OF and NPS specimens collected in parallel from 164 hospitalized patients; 50 bronchoalveolar lavage (BAL) specimens from a subgroup of severe COVID-19 cases were also analysed.
Using Simplexa™ COVID-19 Direct on OF matrix, 100% analytical detection down to 1 TCID50/mL (corresponding to 4 × 10 copies (cp)/mL) was observed. No crossreaction with other viruses transmitted through the respiratory toute was observed. Parallel testing of 337 OF and NPS samples showed highly concordant results (κ = 0.831; 95 % CI = 0.771-0.891), and high correlation of Ct values (r = 0.921; < 0.0001). High concordance and elevated correlation was observed also between OF and BAL. Prolonged viral RNA shedding was observed up to 100 days from symptoms onset (DSO), with 32% and 29% positivity observed in OF and NPS samples, respectively, collected between 60 and 100 DSO.
Simplexa™ COVID-19 Direct assays on OF have high sensitivity and specificity to detect SARS-CoV-2 RNA and provide an alternative to NPS for diagnosis and monitoring SARS-CoV-2 shedding.
鼻咽(NPS)/口咽拭子的 RT-PCR 是诊断 SARS-CoV-2 感染和病毒载量监测的金标准。口腔液(OF)是一种替代的临床样本,易于采集且更安全,可用于 COVID-19 诊断、监测病毒载量和脱落。
对适用于 OF 基质的商业 Simplexa™ COVID-19 直接检测法进行了优化,建立了最佳的检测条件和分析灵敏度。该检测方法用于平行检测了 164 名住院患者的 337 份 OF 和 NPS 标本;还分析了一组严重 COVID-19 病例的 50 份支气管肺泡灌洗液(BAL)标本。
使用 Simplexa™ COVID-19 直接对 OF 基质进行检测,观察到 100%的分析检测下限为 1 TCID50/mL(相当于 4 × 10 拷贝(cp)/mL)。未观察到与通过呼吸道传播的其他病毒发生交叉反应。337 份 OF 和 NPS 样本的平行检测结果高度一致(κ=0.831;95%CI=0.771-0.891),Ct 值相关性高(r=0.921;<0.0001)。OF 与 BAL 之间也观察到了高度一致性和升高的相关性。从症状发作(DSO)开始可观察到长达 100 天的病毒 RNA 持续脱落,在 DSO 后 60 至 100 天采集的 OF 和 NPS 样本中,分别有 32%和 29%呈阳性。
Simplexa™ COVID-19 直接检测 OF 中的 SARS-CoV-2 RNA 具有高灵敏度和特异性,可替代 NPS 用于诊断和监测 SARS-CoV-2 脱落。