Gopalakrishnan Prabhakaran, Jothi Swathi, Lasko Thelma, Denning Stephen
Department of Cardiology, Aultman Hospital, Canton, Ohio.
CASE (Phila). 2020 Jul 16;4(5):439-442. doi: 10.1016/j.case.2020.06.007. eCollection 2020 Oct.
• Probe malfunction during TEE is an underrecognized problem. • MR severity assessment: host (e.g., BP), technical (e.g., gain), probe malfunction. • Clues for malfunction 1: disparity between TTE and TEE; CW and color Doppler. • Clues for malfunction 2: unexplained poor image quality; need for high gain. • Transducer temperature can affect image quality.