Hartnell G G, Kiely E A, Williams G, Gibson R N
Department of Diagnostic Radiology, Royal Postgraduate Medical School, Hammersmith Hospital, London.
Br J Radiol. 1987 Nov;60(719):1063-5. doi: 10.1259/0007-1285-60-719-1063.
The accuracy of different ultrasound methods for determining the volume of urine in the bladder was compared using three methods in current use. Each method was applied to the same ultrasound images from 16 patients with prostatic hypertrophy. The calculated volumes were compared with the true volumes derived by measurement of voided and catheter-drained urine. All methods showed similar degrees of accuracy in quantifying bladder volumes. The range of errors of the best method tested was +/- 35% of true volume and this suggests that ultrasound measurement of bladder volume is not sufficiently accurate for many clinical and research applications.
使用当前使用的三种方法比较了不同超声方法测定膀胱尿量的准确性。每种方法都应用于16例前列腺增生患者的相同超声图像。将计算出的尿量与通过测量排尿量和导尿管引流尿量得出的实际尿量进行比较。所有方法在量化膀胱尿量方面显示出相似的准确度。所测试的最佳方法的误差范围为实际尿量的±35%,这表明对于许多临床和研究应用而言,超声测量膀胱尿量的准确性不足。